Skip to main content

A pelvic kinematic approach for calculating hip angles for active hip disarticulation prosthesis control

Abstract

Background

Control system design for a microprocessor-controlled hip–knee–ankle–foot (HKAF) prosthesis is a challenge since hip disarticulation amputees lack the entire leg and, therefore, only have pelvis movement as user-guided input. This research proposes a method for determining hip joint angles from pelvis movement in a control system for the next generation of powered prostheses.

Method

Three-dimensional pelvic motion and stance time of 10 transfemoral (TF) prosthetic users were used to identify important features and to develop an algorithm to calculate hip angles from pelvis movement based on correlation and linear regression results. The algorithm was then applied to a separate (independent) TF group to validate algorithm performance.

Results

The proposed algorithm calculated viable hip angles during walking by utilizing pelvic rotation, pelvic tilt, and stance time. Small angular differences were found between the algorithm results and motion capture data. The greatest difference was for hip maximum extension angle (2.5 ± 2.0°).

Conclusions

Since differences between algorithm output and motion data were within participant standard deviations, the developed algorithm could be used to determine the desired hip angle from pelvis movements. This study will aid the future development of gait control systems for new active HKAF prostheses.

Introduction

People with an amputation at the hip level have the most difficulty returning to walking [1]. Operating a passive hip disarticulation (HD) prosthesis can be physically demanding, especially for elderly users, and requires sufficient physical fitness [2, 3]. Furthermore, in cases where people continued to use their prostheses, excessive pelvic tilt and rotation during ambulation could eventually lead to spinal injuries [4]. The lack of muscle power at the hip, knee, and ankle/foot can also result in a fixed and slow cadence. A possible solution to reduce physical demand and improve mobility in people with HD amputation is to utilize active actuators to operate the hip joint intelligently. User input is required for hip control; however, identifying viable user input that could be used for prosthetic control is a challenge since hip disarticulation amputees lack the entire hip and leg.

In conjunction with knowledge of 3D gait biomechanics, advancements in actuator and sensor technologies have produced effective control systems for lower extremity prostheses [5]. Powered knee joints that actively aid the prosthetic user in challenging terrains can be beneficial by reducing strain on the body [6, 7]. However, microprocessor-controlled and powered hip joints are only now being investigated in the research domain [8].

Ueyama et al. [8] substituted conventional prosthetic hip and knee joints with conventional fully passive (mechanical) hip joints and semi-active (active dampening) knee joints with robotic motorized joints to actuate the knee and hip. Their prototype used a simple gait control strategy to mirror intact limb motion (echo-controller). A feedback controller allowed the hip to extend while compressing a virtual spring as the participant moved the intact limb forward. The compressed virtual spring's potential energy was released during swing phase, allowing the actuators to swing the prosthesis forward. While this gait control method produced natural gait patterns, the control method was not viable for day-to-day activities due to limited user control and overreliance on the regularity of the steady state gait [9]. Delayed response is another disadvantage of echo controllers [10]. Real-time user control becomes especially crucial while operating a powerful joint capable of propelling the user forward and delayed response of the prosthesis may lead to loss of balance [11]. Mirroring the intact limb is also problematic for stumbles or other asymmetric gait activities (e.g., working in a kitchen, etc.). With the evolution of hip prosthetic actuation, the need for a robust and reliable control mechanism has become essential [12].

Similar to transfemoral (TF), some people with HD amputation are fitted with a prosthesis and receive rehabilitation to use the new prosthesis appropriately [13]. Unlike TF amputees that utilize thigh skeletomuscular segments to help control their prosthesis, HD patients typically rely on their pelvis and torso to operate and control their passive hip–knee–ankle–foot (HKAF) prosthesis [14,15,16]. Since motorized HD prosthesis kinetics and kinematics differ from gait with passive devices [8], applied torque through the motorized hip joint will require a specific gait control strategy [12]. That is, conventional mechanical HD prosthesis biomechanics cannot be directly used for optimal motorized hip joint performance. A more appropriate comparator for walking with a powered hip joint would be TF amputee gait since both powered-hip-HKAF, and TF prostheses can have similar prosthetic knee joints and feet; therefore, a well controlled hip joint could allow an HD amputee to walk as well as a TF amputee [17]. The sole body segment an HD amputee can directly control is the pelvis. Thus, understanding how TF amputees' pelvic motion relates to hip angle during the gait cycle can be the basis of a powered hip joint control approach.

In this research, we identified common pelvic motion features during the gait cycle among TF prosthesis users and determined how these features influence hip flexion and extension. These pelvic features and their relationship with hip rotation were then used to develop an algorithm to estimate the hip angle at each instance in the gait cycle. Thus, this algorithm enables hip angle to be calculated from pelvic motion measurements. Successful evaluation of this algorithm with a separate TF amputee gait dataset demonstrates the viability of this new approach. This pelvis-based approach enables progression to the next phase of powered hip joint control system development, leading to better mobility for people with hip level amputations.

Methods

We investigated TF amputee pelvic motions (pelvic tilt, pelvic obliquity, pelvic rotation) and their relationship with hip rotation. This data was used to develop an algorithm to calculate hip angles throughout the gait cycle, from pelvic motion while walking on a treadmill with no incline. Hip angles calculated by the algorithm were compared with measured (motion-captured) hip angles to quantify algorithm. This study focused on analyzing model performance during steady-state walking.

Databases

Two databases of transfemoral amputee gait were used in this study, for algorithm development and validation. Dataset selection requirements were: all participants must be TF amputees; participants must ambulate without relying on walking assistive devices such as canes, walkers, or treadmill assistive bars; gait data must be from steady-state sequences of level walking tests.

A database from a group of 10 people with unilateral transfemoral amputation with K3 and K4 activity levels [18] was used for algorithm development. Data were collected in the CAREN Extended virtual reality laboratory at The Ottawa Hospital Rehabilitation Centre and processed with C-motion Visual 3D (version 6) using six degrees of freedom model and MATLAB software version 2021a [19,20,21,22]. People walked on a treadmill (level walking) at self-selected walking speeds (1.05 ± 0.27m/s), and ten strides per person were extracted (a total of 100 strides). The development group's mean age was 47 (± 9.4) years, weight was 85 (± 8.6) kg, and height was 176 (± 9) cm. The gait events were processed using foot velocity relative to the pelvis method [23] and through manual verification. This database has been utilized in various biomechanics-related studies published in literature [21, 24, 25]. The Ottawa Health Science Network Research Ethics Board approved the secondary use of this dataset.

Testing group data were obtained from a recently published database ([26, 27]) that used 10 VICON 3D motion capture cameras and a dual belt instrumented Bertec treadmill. Ten people with unilateral transfemoral amputation with K2 and K3 activity levels met our inclusion criteria. Participants walked at different fixed speeds based on their activity level. People with K2 activity levels completed separate walking trials at 0.4, 0.5, 0.6, 0.7, and 0.8 m/s. Those with K3 activity levels walked at 0.6, 0.8, 1.0, 1.2, and 1.4 m/s. The marker data were then processed using modified Vicon plug-in-gait model to obtain the hip and knee kinematics [28, 29]. The pelvic kinematic data were processed using a CODA pelvic model in MATLAB™ R2021a [30]. The testing group's mean age was 47 (± 15) years, weight was 84.5 (± 17.6) kg, and height was 177 (± 11) cm.

Differences between the development and testing groups for activity level, walking speed, or prosthetic components are desirable when evaluating model performance on different groups of participants. This helps to determine model generalizability and avoid cases where better results are due to training and testing on the same group. Prosthesis details for both groups are presented in Tables 1 and 2.

Table 1 Testing group details
Table 2 Development group details

Algorithm development

Pelvic movement in sagittal (anterior/posterior tilt), frontal (lateral tilt or obliquity), and transverse (pelvic rotation) planes were investigated, and gait phase transition timing was assessed to determine their correlation with hip angle throughout the gait cycle.

In this study, any posterior movement of the pelvis and hip joints were considered as positive angular displacement (posterior pelvic tilt, forward pelvic rotation, and hip flexion), while anterior movements were considered as negative angular displacement (anterior pelvic tilt, backward pelvic rotation, and hip extension).

The hip angle calculation algorithm development required:

  1. A.

    Hip angle features: Temporal and spatial hip angle features to calculate hip angular velocity.

  2. B.

    Correlations: Analyze pelvic motion and stance time data to determine common features and correlations with hip angle features.

  3. C.

    Hip angle feature calculation equations: Develop regression equations to calculate the hip angle features using the identified pelvic features in step B.

  4. D.

    Identification of per-person constants: Determine constant parameters unique to each person's gait characteristics that could be used for algorithm development.

  5. E.

    Sequential hip angle calculation: Develop an algorithm for real-time hip angle calculation using parameters obtained in steps C and D.

  6. F.

    Algorithm performance analysis: Evaluate the model with a new data set (testing group).

Each of these will be considered in turn.

  1. A.

    Hip angle features

A typical gait cycle for people with transfemoral amputation includes ([17, 31]): foot contacts the ground (foot strike) with the hip flexed to ~ 30°; hip moves to ~ 5° of extension at the end of stance; hip starts to flex ~ 35° until 80%-90% of stride; and hip flexion decreases ~ 5° in terminal swing to ensure that the prosthetic knee is fully extended before the next foot-strike [32,33,34,35].

Hip angle kinematics can be divided into three periods (Fig. 1), with each period having a linear progression of hip angle by time (i.e., constant slope or constant angular velocity).

  • Period 1 (hip extension): Initiates at foot-strike and ends at 50–60% of the gait cycle. The angular velocity vector is always negative.

  • Period 2 (hip flexion): Initiates at hip max extension and continues through the stance-to-swing transition with minor angular velocity change. The velocity vector is always positive.

  • Period 3 (knee control): Knee joint behavior directly affects hip rotation in this period, with hip rotation acting to keep the prosthetic knee fully extended. For this study, hip angle throughout period 3 was assumed to be constant (angular velocity is zero) until the next foot strike.

Fig. 1
figure 1

Three hip angle periods during gait cycle: hip rotation toward negative angle (extension), hip rotation toward positive angle (flexion), constant angle. Measured curve is the average of 100 strides in the development group

Hip angular velocity was assumed to remain mostly constant throughout each period, simplifying the hip angle calculation. Five features were present for all participants; therefore, these features were used to calculate the constant angular velocity during each period. Spatial features were hip angle at foot strike (HθFS), maximum hip extension angle (HθE), and maximum hip flexion angle (HθF). Temporal features were maximum hip extension time (HτE) and maximum hip flexion time (HτF).

For period 1 (hip extension), the constant angular velocity was calculated using HθFS, HθE, and the difference in time between the two features. For period 2 (hip flexion), the constant angular velocity was calculated using HθE, HθF, and the difference in time between the two features. In period 3 (knee control), hip angle was assumed to be constant at HθF (i.e., zero angular velocity).

The algorithm uses these angular velocity values to calculate hip angle at each time point. Therefore, equations that define the relationships between pelvic kinematics and hip features are required.

  1. B.

    Correlations

Pelvic angular velocity, pelvic tilt, pelvic obliquity, pelvic rotation, and stance time per stride for each person in the development group were analyzed to determine common features that could be used to calculate the constant hip angular displacement and velocities. Twenty-two features were identified as potential candidates. Next, p-earson correlation analyses were applied to each feature candidate to determine which features were most related to the hip angle features. The strongest correlations were (Fig. 2):

  • Pelvic tilt angle at foot strike (PTθ) and hip angle at foot-strike (HθFS): r = 0.95

  • Timing of pelvic rotation angular velocity zero-crossing in early stance (PRZC) and the hip rotation angle range of motion during that period (ΔH): r = -0.75

  • Timing of pelvic tilt angular velocity first zero crossing in midstance (PTZC) and hip max extension time (HτE): r = 0.90

  • PTZC and hip max flexion time (HτF): r = 0.82

  • Stance time (τs) and hip max flexion time (HτF): r = 0.93.

Fig. 2
figure 2

Pelvic feature and stance time correlation with hip angle features. A Pelvic tilt angle at foot strike (PTθ) and hip angle at foot-strike (HθFS), B timing of pelvic angular velocity zero-crossing in early stance (PRZC) and hip angle range of motion during that period (ΔH), C timing of pelvic tilt angular velocity first zero crossing in midstance (PTZC) and hip max extension time (HτE), D timing of pelvic tilt angular velocity first zero crossing in mid-swing (PTZC) and hip max flexion time (HτF), E stance time (τs) and hip max flexion time (HτF)

PTZC (Timing of pelvic rotation angular velocity zero-crossing in early stance) can be detected in real-time by continuously monitoring the 10 most recent pelvic tilt angular velocity data samples. If the most recent sample is positive while the prior 10 samples are negative, positive zero-crossing is detected (Fig. 3).

Fig. 3
figure 3

Pelvic tilt angular displacement (A) and corresponding pelvic tilt angular velocity (B). The red square represents the most recent sample, while the blue stars represent the 10 data samples prior to the most recent sample

PRZC (timing of pelvic tilt angular velocity first zero crossing in mid-swing) is detectable by measuring the pelvic rotation angular velocity zero-crossing (Fig. 4). Initially, angular velocity should be in the positive region. The crossing instance is marked if the angular velocity vector crosses the zero value into the negative angular velocity region.

Fig. 4
figure 4

Pelvic rotation angular displacement (A) and corresponding pelvic rotation angular velocity (B)

  1. C.

    Hip angle feature calculation equations

A series of linear regression equations were determined to calculate the hip angle features based on the linear relationships and correlations between pelvic motion features, gait events, and hip rotation features (Fig. 2). These regression equations are calculated sequentially (from Eqs. 15) as the pelvic features and gait events are detected (Fig. 6).

When foot strike is detected, regression Eq. 1 is used to calculate hip angle.

$${H\theta }_{FS}=24.27+\left(1.45\times PT\theta \right)$$
(1)

where HθFS is hip angle at foot strike and PTθ is pelvic tilt angle at foot strike.

Once pelvic rotation zero crossing is detected (Fig. 4), regression Eq. 2 is used to determines ΔH.

$$\Delta H=-0.6261-\left(59.04\times {PR}_{ZC}\right)$$
(2)

where PRZC is the difference between the time of pelvic rotation angular velocity first zero crossing and foot-strike time. ΔH is the difference between hip angle at foot-strike and hip angle at PRZC time.

Approaching the end of stance phase, pelvic tilt zero crossing is detected (Fig. 3) and used to calculate hip max extension time using regression Eq. 3 and hip max flexion time using Eq. 4.

$${H\uptau }_{E}={0.0462+(1.15\times PT}_{ZC})$$
(3)
$${H\uptau }_{F\alpha }=0.3461 +\left(1.35\times {PT}_{ZC}\right)$$
(4)

where PTZC is the time of pelvic tilt angular velocity zero crossing, HτE is the hip max extension time, and Hτ is the hip max flexion time at the PTZC instant.

During the foot-off instance, HτF is recalculated again in Eq. 5 to take advantage of the higher HτF and τS correlation value (R = 0.94) compared to HτF and PTZC (R = 0.84).

$${H\uptau }_{F\beta }=0.0874 +\left(1.18\times {\uptau }_{S}\right)$$
(5)

where τS is the gait stance time, and Hτ is max hip flexion time at foot-off.

  1. D.

    Per person, unique constants

For hip angle calculation using the developed algorithm, two unique input constants were defined for each person after analyzing the development dataset: mean hip max extension and mean hip max flexion angles. In practice, during the fitting process, a prosthetist would increase or decrease these constants for each patient until results are satisfactory [36]. That is, iterative changes are made to the values during the prosthetic fitting process until gait is acceptable to the end-user and clinician, similar to the process for tuning other microprocessor controlled prosthetic joints.

Hip max extension and max flexion angles tended to be in a unique but limited range for each TF amputee in the development group. Studies suggested that this unique hip ROM is due to prosthesis user control strategies to maintain balance during walking [37, 38]. The per-person-specific hip max extension and max flexion angle constants for both development and testing groups are provided in Table 3.

Table 3 Per-person unique constants of development and testing group datasets
  1. E.

    Hip angle calculation

Figure 5 shows the general block diagram of the hip angle calculation algorithm for steady state gait. Initially, pelvic features are extracted from pelvic motion data. Then, using the hip angle feature calculation equations and foot-off time, the hip parameters necessary for constant hip angular velocity calculations are obtained. In the algorithm's last two stages, the calculated hip features are used to calculate hip angular velocity and hip angle.

Fig. 5
figure 5

Gait cycle hip angle calculation block diagram

The six sequences shown in Fig. 6 represent five different hip angle equations used in the algorithm. Since the algorithm was developed to run in real-time, pelvic features are detected by the algorithm at different times during the gait cycle (since people walk at different speeds). Therefore, different hip calculation equations are used depending on the timing of each pelvic feature detection. The hip extension angle (period 1) is calculated during the first three sequences. During sequences 4 and 5, the hip angle during hip flexion (period 2) is calculated, and during sequence 6 the hip angle during period 3 is calculated. For detailed mathematics of hip calculations for each sequence see Additional file 1.

Fig. 6
figure 6

Calculated hip angle throughout the gait cycle. PRZC: pelvic rotation zero-crossing time, PTZC: pelvic tilt zero-crossing time, HτE: hip max extension time, τs: stance time, HτF: hip max flexion time. Sequences are foot-strike to PRZC (sequence 1, solid red line), PRZC to PTZC (sequence 2, solid black line), PTZC to HτE (sequence 3, solid blue line), HτE to τFO (sequence 4, red dashed line), τFO to HτF (sequence 5, blue dashed line), and HτF to the end of the gait cycle (sequence 6, black dashed line)

  1. F.

    Algorithm evaluation

The developed algorithm was applied to both testing and development groups. Algorithm performance was assessed by calculating hip angle feature differences between calculated (algorithm) and motion capture system data (ground truth). To provide a reference for whether the differences were appropriate or beyond typical participant variability, the calculated differences were compared with participant averaged hip feature standard deviations in the testing and development set.

Results

Development group

The root-mean-square-error (RMSE) between the measured and calculated hip trajectory and the average differences between motion-captured features and algorithm calculated data for the development group are shown in Table 4. Participants walked at self-paced speeds ranging from 0.67 m/s to 1.55 m/s, with a mean of 1.05 ± 0.27 m/s.

Table 4 Development group average differences and standard deviations between measured and calculated features for all participants, average correlation between calculated and measured features, and average root-mean-squared-error between calculated and measured hip trajectories

Hip max extension angle had the largest difference compared to other features, with the per-participant averaged difference ranging from 1.5° to 3.5°. The greatest overall mean difference was for hip max extension angle (2.4 ± 1.8°). Hip temporal features (HτE and HτF) had the lowest difference, indicating that pelvic motion temporal features were in sync with hip temporal features during the gait cycles. The calculated hip features closely agreed with the measured data, with correlation values varying from 0.94 to 0.99. Similarly, the average RMSE between the measured and calculated hip trajectory for development group was small (3.71 ± 1.13°) relative to average hip range of motion (44.05 ± 7.38°).

The mean and standard deviation of hip feature angles for each participant in the development group are provided in Table 5.

On average, standard deviations were 1.7° at foot-strike, 1.4° at max extension, and 2.0° at max flexion (Table 5). The mean differences in Table 4 were compared with the average standard deviation for each participant in Table 5. The algorithm calculated features were at most 1° greater than the average standard deviations for development group participants walking at self-paced speed.

Table 5 Motion captured hip angle (average and standard deviation) of development group walking at a self-paced speed

Testing group

The algorithm was applied to 2403 strides from the testing group. The outcomes (Table 6) showed small differences between measured and algorithm calculated data. Similar to the development group results (Table 4), mean hip max extension angle had the largest difference (3.4 ± 2.8°). The average RMSE comparison between participants revealed that Participant 7 exhibited the greatest deviation from the other participants in the testing group while having the lowest correlation. All per-participant correlations were above 0.94. Per-feature correlation analysis demonstrated that the lowest correlation was for HθE (0.79). The average RMSE of measured and algorithm calculated hip trajectories in the testing group was small (5.01 ± 2.01°) compared to the hip range of motion (43.42 ± 8.18°), but larger than the development group.

Table 6 Testing group average per-feature differences and standard deviations for all participants, average correlation between calculated and measured features, and average root-mean-squared-error between calculated and measured hip trajectories

Measured hip features for the testing group are provided in Table 7. Although testing group participants walked at a fixed speed, the hip feature standard deviations were approximately 0.3° more than the development group results. On average, hip feature standard deviations for the testing group were 2.6° at foot-strike, 3.0° at max extension, and 1.8° at max flexion. The measured and calculated hip feature differences in Table 6 were compared with the mean hip feature standard deviations in Table 7. Algorithm-calculated hip angle at foot-strike (2.0°) and hip max flexion angle (1.9°) were within the standard deviations for each feature. However, hip max extension angle difference was overestimated by 1.2°.

Table 7 Motion captured hip angle (average and standard deviation) of testing group

Discussion

In this research, we investigated pelvic movements of transfemoral amputees and developed a viable algorithm for calculating hip angles based on pelvic motion and stance time. With the introduction of powered prosthetic joints, the limitations of existing prostheses are mitigated by transferring a large amount of the mechanical work to the actuator, reducing the reliance on the user's physical capabilities to operate the prosthesis [8, 39]. Since only the pelvis segment is available for user control of a motorized hip joint, a new approach is required to convert pelvic motion (system input) into a hip angular map for a gait controller to follow. The approach proposed in this research addresses this need.

The newly developed algorithm requires only three features (pelvic tilt angle at foot strike, timing of pelvic rotation angular velocity zero-crossing in early stance, and timing of pelvic tilt angular velocity first zero crossing in midstance) and stance time data for hip angle calculation. After pelvic angle analysis, pelvic tilt and rotation exhibited common behaviours among most people in the development and testing groups and hence could be used for hip angle calculation. Pelvic angle features that correlated well with the hip angle features were then extracted for algorithm development.

Pelvic obliquity varied greatly between strides in both groups and although some common features were found, none were viable features for a motorized hip prosthetic control system. Although the pelvic tilt zero-crossing feature (PTZC) was common among all persons in both groups, pelvic tilt of participant 7 in the testing group tended to be irregular relative to other people, even though there was no gait pathology observed during data acquisition. This irregularity resulted in many pelvic feature detection errors. Adequate gait training is required to avoid pelvic motions leading to pelvic feature or hip angle errors.

Model output from both development and testing datasets had very high correlations between motion capture and algorithm calculated data (R = 0.99 for development group and R = 0.95 for testing group). The high correlations indicated that the assumed three-period hip angle model (Fig. 1) closely matched typical hip angles trends for transfemoral amputees. The low RMSE hip trajectory indicated that the calculated temporal and special features closely matched the measured data. The lowest correlation observed in the testing group results in Table 6 belonged to HθE (0.79) and was likely due to compensatory kinematic adjustments of participants to maintain a consistent speed while walking on a fixed-speed treadmill [40].

Motion capture and algorithm-calculated hip angle features had small differences for all testing and development group participants. The largest difference was hip max extension angle for the development group (2.4 ± 1.8°) and hip max extension angle for the testing group (3.4 ± 2.8°). Furthermore, the temporal hip feature differences (hip max extension time, max flexion time) were close to zero, indicating that the algorithm can accurately calculate temporal hip features. Comparing the differences with standard deviations of both datasets showed that only the calculated hip max extension angle exceeded the participant's typical standard deviation (by 1° for development group and by 1.2° for testing group). All other hip feature differences between calculated and measured values were within the measured standard deviations. We hypothesize that, in practice, exceeding the typical hip angles by 1.2° should not substantially affect an active hip prosthesis user's stability or comfort. However, further testing with a prototype prosthesis is necessary to assess the effect of the hip angle algorithm on mobility.

Analysis of development and testing group data showed that hip max extension and hip max flexion angles during the gait cycle tended to vary slightly in a unique range for each person. Thus, each participant's hip max extension and flexion angles averaged over all strides were used as input constants for each participant in this study (Fig. 5). As a person walks, hip max extension and flexion angles remain consistent or vary slightly within a limited range, even with changes in walking speed. It is important to note that per-person unique constants were determined because people in both groups walked at speeds between 0.4 m/s to 1.5 m/s; therefore, an additional analysis could be required for people with preferred walking speeds outside this range. When the algorithm is used in practice for a powered hip control system, the clinician responsible for prosthetic fitting should adjust the hip max extension angle and hip max flexion angle constants to accommodate the user. These parameters would then be entered into the system as unique per-person constants.

Our feature and event-based approach has two key advantages over the echo controller approach proposed by Ueyama et al. [8]: our approach allows for gait speed adaptation, and also enables real-time control. These benefits are particularly valuable for prosthetic users who need to operate their prosthesis in environments where steady-state gait is not feasible. Echo controllers rely on the motion of the intact limb to determine joint trajectory and gait speed, which means it will take twice as long and require twice the distance to change gait speed [41]. By contrast, our proposed feature and event-based trajectory adjustment approach continuously adjusts the temporal and spatial aspects of hip kinematics in real-time, based on the user's response, which can be detected by moving the pelvis, applying foot-strike, and performing foot-off. This means prosthetic users can adjust their gait speed in real time without relying on the intact limb. Consequently, prosthetic users will be better equipped to navigate confined spaces or crowded environments where repeated and rapid speed adjustments are necessary. A desirable microprocessor-controlled prosthesis should seamlessly integrate all sensors within the prosthesis, allowing the user to simply charge and wear the device. However, most prostheses with echo controller integration are dependent on sensors located on the intact side and have not garnered positive feedback from users [9].

While we utilized correlation analysis to identify the relationship between pelvic motions and sagittal hip angles, the calculated hip rotation velocities based on these correlations were linear for all three periods (Fig. 1). Finite state impedance controller (virtual springs and dampers between thighs and pelvis) and virtual holonomic constraints can introduce angular velocity non-linearities that more closely resemble the biological kinematics [17, 34, 42]. Therefore, for future studies on active HKAF prosthesis control strategies, integration of finite state impedance controller or virtual holonomic constraints, in addition to our algorithm may be necessary to achieve appropriate hip kinematics. Additionally, bio signal analysis tools, such as wavelet analysis [43], and biomechanical pattern recognition approaches, such as machine learning and state estimation [44], could potentially be used to identify more pelvic motion features for a more accurate hip trajectory predication.

This research used gold-standard 3D motion capture data for pelvic kinematics measurements [45]. However, wearable systems such as lower limb prosthetics need integrated measurement technology to provide real-time input for prosthesis control. Recent studies have investigated the feasibility of using inertial measurement units (IMUs) to measure lower extremity kinematics [46, 47]. In a study by Berner et al. [46], multiple IMUs were used to measure the kinematics of different body segments with high accuracy. In another study, researchers accurately measured pelvic kinematics of an able-body participant by utilizing a single IMU mounted to the lower back [47]. In theory, the required pelvic motion input data for the developed algorithm could be provided by an IMU integrated into the socket connection component (i.e., IMU moves with the socket, which is attached to the pelvis, to provide pelvis kinematics) and a loadcell in the joint to measure applied forces such as at foot strike.

Considering the greater reliance of HD amputees on pelvic and trunk motions, further algorithm tuning of the pelvis feature detection and hip angle calculation algorithm are anticipated to optimize the prosthesis gait controller. By utilizing powered and intelligent hip joints, gait in people with an HD amputation could be improved to a level similar to people with transfemoral amputation. As powered hip prostheses become more widely available, new gait data could be collected with HD participants and used to improve the algorithm.

Ultimately, the algorithm proposed in this research is an important step towards creating a powered hip prosthesis controller that enables more efficient movement in the person's chosen environment, all with less strain on the lower back and less effort.

Limitations

The testing group people walked on a treadmill at fixed speeds that were different than the development group. Furthermore, all gait cycle data used in this study were at a steady state. Therefore, it is possible that steady state walking reduced the natural gait variation and thus contributed to the high accuracy of our algorithm.

Conclusion

A viable algorithm was developed and evaluated to calculate hip angles throughout the gait cycle using only pelvis angles and stance time data. This new hip angle calculation method lays the groundwork for the future development of an intelligent powered hip prosthesis control system since it permits the use of measurable pelvic data to generate hip data as user input for a person's HKAF prosthesis. Based on our gait kinematics and gait event time analysis, hip angle calculations would require sensors in the prosthesis capable of detecting pelvic kinematics and stance phase.

Availability of data and materials

The dataset acquired from Ottawa Hospital Research Institute is not authorized for external use. The dataset provided by S. Hood is available via scientific data database [26].

Abbreviations

HD:

Hip disarticulation

TF:

Transfemoral

HKAF:

Hip–Knee–Ankle–Foot

RMSE:

Root-mean-square-error

FS :

Hip angle at foot-strike instant

E :

Maximum hip extension angle

F :

Maximum hip flexion angle

E :

Maximum hip extension time

F :

Maximum hip flexion time

\({{\overline{\text{H}\theta}}_{\text{E}}}\) :

Hip max extension angle constant

\({{\overline{\text{H}\theta}}_{\text{F}}}\) :

Hip max flexion angle constant

PTθ:

Pelvic tilt angle at foot strike

PRZC :

Timing of pelvic rotation angular velocity zero-crossing

ΔH:

Hip rotation angle range of motion during pelvic rotation angular velocity zero-crossing

PTZC :

Pelvic tilt angular velocity first zero crossing

Τs:

Stance time

PNO:

Participant number

Corr:

Pearson correlation

References

  1. Fernández A, Formigo J. Are Canadian prostheses used? A long-term experience. Prosthet Orthot Int. 2005;29:177–81.

    Article  PubMed  Google Scholar 

  2. Chin T, Oyabu H, Maeda Y, Takase I, Machida K. Energy consumption during prosthetic walking and wheelchair locomotion by elderly hip disarticulation amputees. Am J Phys Med Rehabil. 2009;88(5):399–403. https://0-doi-org.brum.beds.ac.uk/10.1097/PHM.0b013e3181a0dbe2.

    Article  PubMed  Google Scholar 

  3. Chin T, Kuroda R, Akisue T, Iguchi T, Kurosaka M. Energy consumption during prosthetic walking and physical fitness in older hip disarticulation amputees. J Rehabi Res Dev. 2012;49:1255–60.

    Article  Google Scholar 

  4. Ludwigs E, Bellmann M, Schmalz T, Blumentritt S. Biomechanical differences between two exoprosthetic hip joint systems during level walking. Prosthet Orthot Int. 2010;34(4):449–60. https://0-doi-org.brum.beds.ac.uk/10.3109/03093646.2010.499551.

    Article  PubMed  Google Scholar 

  5. Boone D. Prosthetists and orthotists: an evolution from mechanic to clinician. Prosthet Orthot Int. 2020;44:368–72.

    Article  PubMed  Google Scholar 

  6. Brandt A, Wen Y, Liu M, Stallings J, Huang HH. Interactions between transfemoral amputees and a powered knee prosthesis during load carriage. Sci Rep. 2017;7(1):14480. https://0-doi-org.brum.beds.ac.uk/10.1038/s41598-017-14834-7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  7. Martinez-villalpando EC, Herr H. Agonist-antagonist active knee prosthesis: a preliminary study in level-ground walking. J Rehabil Res Dev. 2009;46:361–73.

    Article  PubMed  Google Scholar 

  8. Ueyama Y, Kubo T, Shibata M. Robotic hip-disarticulation prosthesis: evaluation of prosthetic gaits in a non-amputee individual. Adv Robot. 2019;34(1):37–44. https://0-doi-org.brum.beds.ac.uk/10.1080/01691864.2019.1705908.

    Article  Google Scholar 

  9. Li L, Wang X, Meng Q, Chen C, Sun J, Yu H. Intelligent knee prostheses: a systematic review of control strategies. J Bionic Eng. 2022;19:1242.

    Article  Google Scholar 

  10. Varol HA, Goldfarb M. Real-time Intent Recognition for a Powered Knee and Ankle Transfemoral Prosthesis. IEEE.

  11. Ferreira C, Reis LP, Santos CP. Review of Control Strategies for Lower Limb Prostheses. Robot 2015: Second Iberian Robotics Conference. Advances in Intelligent Systems and Computing: Springer International Publishing; 2016. p. 209-20

  12. Tucker MR, Olivier J, Pagel A, Bleuler H, Bouri M, Lambercy O, et al. Control strategies for active lower extremity prosthetics and orthotics: a review. J Neuroeng Rehabil. 2015;12(1):1. https://0-doi-org.brum.beds.ac.uk/10.1186/1743-0003-12-1.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Edelstein JE. Amputations. In: Timothy LK, Ron S, John OB, Michael LM, editors. Comprehensive guide to geriatric rehabilitation. 8th ed. Oxford: Churchill Livingstone; 2014. p. 337–40.

    Chapter  Google Scholar 

  14. Baum BS, Schnall BL, Tis JE, Lipton JS. Correlation of residual limb length and gait parameters in amputees. Injury. 2008;39(7):728–33. https://0-doi-org.brum.beds.ac.uk/10.1016/j.injury.2007.11.021.

    Article  PubMed  Google Scholar 

  15. Engstrom B, Van de Ven C. Therapy for amputees. 3rd ed. Edinburgh: Churchill Livingstone; 1999.

    Google Scholar 

  16. Devan H, Carman A, Hendrick P, Hale L, Ribeiro DC. Spinal, pelvic, and hip movement asymmetries in people with lower-limb amputation: systematic review. J Rehabil Res Dev. 2015;52:1–19.

    Article  PubMed  Google Scholar 

  17. Eslamy M, Oswald F, Schilling AF. Estimation of Knee Angles Based on Thigh Motion: A Functional Approach and Implications for High-Level Controlling of Active Prosthetic Knees. IEEE Control Systems: Institute of Electrical and Electronics Engineers Inc.; 2020. p. 49–61.

  18. Gailey RS, Roach KE, Applegate EB, Cho B, Cunniffe B, Licht S, et al. The amputee mobility predictor: an instrument to assess determinants of the lower-limb amputee’s ability to ambulate. Arch Phys Med Rehabil. 2002;83(5):613–27. https://0-doi-org.brum.beds.ac.uk/10.1053/apmr.2002.32309.

    Article  PubMed  Google Scholar 

  19. Wu G, Siegler S, Allard P, Kirtley C, Leardini A, Rosenbaum D, et al. ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion–part I: ankle, hip, and spine. Int Soc Biomech J Biomech. 2002;35(4):543–8. https://0-doi-org.brum.beds.ac.uk/10.1016/s0021-9290(01)00222-6.

    Article  Google Scholar 

  20. Wu G, van der Helm FC, Veeger HE, Makhsous M, Van Roy P, Anglin C, et al. ISB recommendation on definitions of joint coordinate systems of various joints for the reporting of human joint motion–Part II: shoulder, elbow, wrist and hand. J Biomech. 2005;38(5):981–92. https://0-doi-org.brum.beds.ac.uk/10.1016/j.jbiomech.2004.05.042.

    Article  CAS  PubMed  Google Scholar 

  21. Sturk JA, Lemaire ED, Sinitski EH, Dudek NL, Besemann M, Hebert JS, et al. Maintaining stable transfemoral amputee gait on level, sloped and simulated uneven conditions in a virtual environment. Disabil Rehabil Assist Technol. 2019;14:226–35.

    Article  PubMed  Google Scholar 

  22. Sinitski EH, Lemaire ED, Baddour N. Evaluation of motion platform embedded with force plate-instrumented treadmill. J Rehabil Res Dev. 2015;52:221–34.

    Article  PubMed  Google Scholar 

  23. Zeni JA Jr, Richards JG, Higginson JS. Two simple methods for determining gait events during treadmill and overground walking using kinematic data. Gait Posture. 2008;27(4):710–4. https://0-doi-org.brum.beds.ac.uk/10.1016/j.gaitpost.2007.07.007.

    Article  PubMed  Google Scholar 

  24. Sturk JA, Lemaire ED, Sinitski E, Dudek NL, Besemann M, Hebert JS, et al. Gait differences between K3 and K4 persons with transfemoral amputation across level and non-level walking conditions. Prosthet Orthot Int. 2018;42(6):626–35. https://0-doi-org.brum.beds.ac.uk/10.1177/0309364618785724.

    Article  PubMed  Google Scholar 

  25. Kendell C, Lemaire ED, Kofman J, Dudek N. Gait adaptations of transfemoral prosthesis users across multiple walking tasks. Prosthet Orthot Int. 2016;40(1):89–95. https://0-doi-org.brum.beds.ac.uk/10.1177/0309364614568410.

    Article  PubMed  Google Scholar 

  26. Hood S, Ishmael MK, Gunnell A, Foreman KB, Lenzi T. A kinematic and kinetic dataset of 18 above-knee amputees walking at various speeds. Sci Data. 2020;7:150.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Hood S, Lenzi T. Lower Limb Kinetic and Kinematic Data of 18 Above Knee Amputees. 2020.

  28. Vicon. Plug-in gait reference guide. In: Limited VMS, editor.2021.

  29. Collins TD, Ghoussayni SN, Ewins DJ, Kent JA. A six degrees-of-freedom marker set for gait analysis: repeatability and comparison with a modified Helen Hayes set. Gait Posture. 2009;30(2):173–80. https://doi.org/10.1016/j.gaitpost.2009.04.004.

    Article  PubMed  Google Scholar 

  30. O’Sullivan K, Clifford A, Hughes L. The reliability of the CODA motion analysis system for lumbar spine analysis: a pilot study. Physiother Pract Res. 2010;31(1):16–22. https://0-doi-org.brum.beds.ac.uk/10.3233/ppr-2010-31104.

    Article  Google Scholar 

  31. Eberly VJ, Mulroy SJ, Gronley JK, Perry J, Yule WJ, Burnfield JM. Impact of a stance phase microprocessorcontrolled knee prosthesis on level walking in lower functioning individuals with a transfemoral amputation. Prosthet Orthot Int. 2014;38:447–55.

    Article  PubMed  Google Scholar 

  32. Jaegers SM, Arendzen JH, de Jongh HJ. Prosthetic gait of unilateral transfemoral amputees: a kinematic study. Arch Phys Med Rehabil. 1995;76(8):736–43. https://0-doi-org.brum.beds.ac.uk/10.1016/s0003-9993(95)80528-1.

    Article  CAS  PubMed  Google Scholar 

  33. Shandiz MA, Farahmand F, Osman NAA, Zohoor H. A robotic model of transfemoral amputee locomotion for design optimization of knee controllers. Int J Adv Robot Syst. 2013;10:161.

    Article  Google Scholar 

  34. Sup F, Bohara A, Goldfarb M. Design and control of a powered transfemoral prosthesis. Int J Rob Res. 2008;27(2):263–73. https://0-doi-org.brum.beds.ac.uk/10.1177/0278364907084588.

    Article  PubMed  PubMed Central  Google Scholar 

  35. Segal AD, Orendurff MS, Klute GK, McDowell ML, Pecoraro JA, Shofer J, et al. Kinematic and kinetic comparisons of transfemoral amputee gait using C-Leg and Mauch SNS prosthetic knees. J Rehabil Res Dev. 2006;43(7):857–70. https://0-doi-org.brum.beds.ac.uk/10.1682/jrrd.2005.09.0147.

    Article  PubMed  Google Scholar 

  36. Edelstein JE. Amputations and prostheses. In: Cameron MH, Monroe LG, editors. Physical rehabilitation. Saint Louis: W.B. Saunders; 2007. p. 267–99.

    Chapter  Google Scholar 

  37. Koehler-McNicholas SR, Lipschutz RD, Gard SA. The biomechanical response of persons with transfemoral amputation to variations in prosthetic knee alignment during level walking. J Rehabil Res Dev Serv. 2016;53:1089–106.

    Article  Google Scholar 

  38. Heitzmann DWW, Leboucher J, Block J, Gunther M, Putz C, Gotze M, et al. The influence of hip muscle strength on gait in individuals with a unilateral transfemoral amputation. PLoS ONE. 2020;15(9): e0238093. https://0-doi-org.brum.beds.ac.uk/10.1371/journal.pone.0238093.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  39. Blumentritt S, Ludwigs E, Bellmann M, Boiten H. The New Helix 3D Hip Joint. 2008.

  40. Sinitski EH, Lemaire ED, Baddour N, Besemann M, Dudek NL, Hebert JS. Fixed and self-paced treadmill walking for able-bodied and transtibial amputees in a multi-terrain virtual environment. Gait Posture. 2015;41(2):568–73. https://0-doi-org.brum.beds.ac.uk/10.1016/j.gaitpost.2014.12.016.

    Article  PubMed  Google Scholar 

  41. Vallery H, Burgkart R, Hartmann C, Mitternacht J, Riener R, Buss M. Complementary limb motion estimation for the control of active knee prostheses. Biomed Tech (Berl). 2011;56(1):45–51. https://0-doi-org.brum.beds.ac.uk/10.1515/BMT.2010.057.

    Article  PubMed  Google Scholar 

  42. Horn JC, Gregg RD. Nonholonomic virtual constraints for control of powered prostheses across walking speeds. IEEE Trans Control Syst Technol. 2022;30(5):2062–71. https://0-doi-org.brum.beds.ac.uk/10.1109/tcst.2021.3133823.

    Article  PubMed  Google Scholar 

  43. Akay M. Wavelets in biomedical engineering. Ann Biomed Eng. 1995;23(5):531–42. https://0-doi-org.brum.beds.ac.uk/10.1007/BF02584453.

    Article  CAS  PubMed  Google Scholar 

  44. Chalvatzaki G, Papageorgiou XS, Maragos P, Tzafestas CS. User-Adaptive Human-Robot Formation Control for an Intelligent Robotic Walker Using Augmented Human State Estimation and Pathological Gait Characterization. IEEE.

  45. Ferrari A, Benedetti MG, Pavan E, Frigo C, Bettinelli D, Rabuffetti M, et al. Quantitative comparison of five current protocols in gait analysis. Gait Posture. 2008;28(2):207–16. https://0-doi-org.brum.beds.ac.uk/10.1016/j.gaitpost.2007.11.009.

    Article  PubMed  Google Scholar 

  46. Berner K, Cockcroft J, Morris LD, Louw Q. Concurrent validity and within-session reliability of gait kinematics measured using an inertial motion capture system with repeated calibration. J Bodyw Mov Ther. 2020;24(4):251–60. https://0-doi-org.brum.beds.ac.uk/10.1016/j.jbmt.2020.06.008.

    Article  PubMed  Google Scholar 

  47. Bolink SA, Naisas H, Senden R, Essers H, Heyligers IC, Meijer K, et al. Validity of an inertial measurement unit to assess pelvic orientation angles during gait, sit-stand transfers and step-up transfers: Comparison with an optoelectronic motion capture system. Med Eng Phys. 2016;38(3):225–31. https://0-doi-org.brum.beds.ac.uk/10.1016/j.medengphy.2015.11.009.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This research was funded by Mitacs (IT17303). However, the funding body had no role in the collection, analysis and interpretation of data or in writing the manuscript.

Author information

Authors and Affiliations

Authors

Contributions

EDL and NB coordinated the study. FG conducted data analysis and interpretation. All authors participated in the drafting, editing and review of the final manuscript.

Corresponding author

Correspondence to Farshad Golshan.

Ethics declarations

Ethics approval and consent to participate

The secondary usage of participants' data of development group dataset was approved by Ottawa Hospital Research Institute (OHRI) ethics board. Ethics approval of testing group dataset is not applicable. The data was provided online for public use.

Consent for publication

Not applicable.

Competing interests

The authors declare that there are no conflicts of interest related to this research or this manuscript.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1.

Real-time hip angle calculation algorithm

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Golshan, F., Baddour, N., Gholizadeh, H. et al. A pelvic kinematic approach for calculating hip angles for active hip disarticulation prosthesis control. J NeuroEngineering Rehabil 20, 152 (2023). https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-023-01273-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-023-01273-x