Skip to main content

Table 3 Reliability analysis: test–retest absolute agreement

From: A portable system to measure knee extensor spasticity after spinal cord injury

 

ICC (2,1)

95% CI

F

dof

p-value

Control

 OTS

0.876

0.775–0.933

16.5

42

< 0.001

 IMU

0.864

0.763–0.924

13.5

42

< 0.001

SCI

 OTS

0.932

0.874–0.964

27.9

37

< 0.001

 IMU

0.925

0.861–0.960

25.3

37

< 0.001

Overall

 OTS

0.952

0.927–0.969

40.4

80

< 0.001

 IMU

0.951

0.925–0.968

39.8

80

< 0.001

  1. ICC intraclass correlation coefficient, OTS optical tracking system, IMU  inertia motion unit based system, CI confidence interval, dof degrees of freedom