Usability and performance validation of an ultra-lightweight and versatile untethered robotic ankle exoskeleton

Abstract

Background

Ankle exoskeletons can improve walking mechanics and energetics, but few untethered devices have demonstrated improved performance and usability across a wide range of users and terrains. Our goal was to design and validate a lightweight untethered ankle exoskeleton that was effective across moderate-to-high intensity ambulation in children through adults with and without walking impairment.

Methods

Following benchtop validation of custom hardware, we assessed the group-level improvements in walking economy while wearing the device in a diverse unimpaired cohort (n = 6, body mass = 42–92 kg). We also conducted a maximal exertion experiment on a stair stepping machine in a small cohort of individuals with cerebral palsy (CP, n = 5, age = 11–33 years, GMFCS I-III, body mass = 40–71 kg). Device usability metrics (device don and setup times and System Usability Score) were assessed in both cohorts.

Results

There was a 9.9 ± 2.6% (p = 0.012, range = 0–18%) reduction in metabolic power during exoskeleton-assisted inclined walking compared to no device in the unimpaired cohort. The cohort with CP was able to ascend 38.4 ± 23.6% (p = 0.013, range = 3–132%) more floors compared to no device without increasing metabolic power (p = 0.49) or perceived exertion (p = 0.50). Users with CP had mean device don and setup times of 3.5 ± 0.7 min and 28 ± 6 s, respectively. Unimpaired users had a mean don time of 1.5 ± 0.2 min and setup time of 14 ± 1 s. The average exoskeleton score on the System Usability Scale was 81.8 ± 8.4 (“excellent”).

Conclusions

Our battery-powered ankle exoskeleton was easy to use for our participants, with initial evidence supporting effectiveness across different terrains for unimpaired adults, and children and adults with CP.

Trial registration Prospectively registered at ClinicalTrials.gov (NCT04119063) on October 8, 2019.

Background

Ankle exoskeletons hold potential to augment walking performance in unimpaired individuals and in individuals with neurological conditions [1,2,3,4]. The ankle joint is a frequent target for powered assistance due to its critical role in efficient bipedal locomotion [5,6,7] and because it is a commonly affected joint in individuals with neurological deficits [8, 9]. Individuals with cerebral palsy (CP), for example, typically have ankle plantarflexor weakness and limited push-off power that contributes to slow, inefficient walking, particularly on graded terrain, like stairs [10,11,12].

Unburdened by the need to carry motors and a power supply, users walking with tethered ankle plantarflexor assistance have consistently demonstrated improved walking economy for nearly a decade [7, 13,14,15]. However, achieving improvements in walking economy with untethered ankle exoskeletons has apparently been more challenging, with only a small number of studies reporting activity performance benefits compared to walking without the device [1, 3, 16,17,18]. Untethered ankle exoskeletons capable of mobility augmentation outside of the laboratory follow two general design approaches: placing motors on the shank close to the joint or placing motors at the waist. Opting to minimize mass and the physical profile added to the lower-limb, Awad et al. [1, 17] developed a soft exosuit with waist-mounted motors that improved paretic limb function, walking speed and walking economy in stroke survivors. Mooney et al. [3] took a shank mounted motor approach instead, and addressed the metabolic detriment of adding mass distally on the leg by incorporating a clever mechanical design achieving high torque and power output, and demonstrated improvements in loaded and unloaded walking in healthy adults; this appears to be the only published work demonstrating a group-level improvement in energy efficiency in unimpaired individuals when walking with an untethered, battery-powered ankle exoskeleton compared to no device.

For several years, our group has worked on untethered, low-torque ankle exoskeletons for children and young adults with CP. We have demonstrated that bilateral assistance proportional to the user’s biological ankle moment during stance phase [19, 20] can improve key metrics such as energy expenditure and walking speed in small cohorts with CP during level walking [18, 21]. However, early prototypes had poor reliability and durability, and proved ineffective for individuals of body mass greater than approximately 45 kg because of limited torque production and significant motion of the ankle assembly relative to the shank and foot. Additionally, these prior exoskeletons were cumbersome to don and doff, designed without consideration for usability, and control was limited to a computer-based researcher interface. Usability factors are important yet under-researched aspects of wearable lower-limb exoskeleton design that hold practical implications for real-world deployment. Devices intended to augment mobility in the community should be easy to don and operate, with portable and intuitive user interfaces. The ability of individuals with CP to put on and operate an ankle exoskeleton without researcher or technician intervention remains unknown.

The first goal of this study was to design a novel cable-driven ankle exoskeleton, validate custom torque and angle sensors, and evaluate electromechanical performance during ambulation (Fig. 1). Our second goal was to highlight the relevance and versatility of this device by demonstrating its ability to reduce the energy cost of fast incline walking in healthy adults, and on distance achieved during a maximal exertion stair-stepping exercise in CP. We selected these moderate- to high-intensity activities for these human performance experiments because we believe such activities reflect the real utility of ankle exoskeletons in both unimpaired and impaired populations, namely, augmenting ambulatory activities that have elevated ankle plantarflexor demand. We hypothesized both cohorts would have significant improvements while walking with vs without the device. Our final objective was to complete a usability assessment, quantifying the time for users or their caretakers, if applicable, to don and set up the device without researcher intervention. We hypothesized that individuals could don, calibrate, and receive assistance from the device in less than 5 minutes.

Fig. 1
figure 1

Exoskeleton mass breakdown, exoskeleton control overview, and protocol summary. A Device components and mass. Values are mass per leg except for the waist assembly. B High- and low-level control layers. The high-level controller was responsible for gait event detection (i.e., toe-off and heel strike) and assistive torque profile generation. During stance, a forefoot force sensor signal was an input to a proportional joint moment controller (PJMC) that generated an adaptive plantarflexor torque profile in real time [17]. During swing, a constant dorsiflexor torque was prescribed. The low-level controller tracked the plantarflexor and dorsiflexor torque profiles using a closed-loop PD controller. C A summary of the experiments, cohorts, conditions, and measurements analyzed in this study

Methods

Exoskeleton design

We designed a lightweight bilateral, bidirectional battery-powered ankle exoskeleton (Fig. 1AB). Waist mounted motors actuated a pulley assembly at the ankle via a chain-to-cable transmission system. The instrumented ankle joint pulley was mounted within a carbon fiber tube that also supported cable housing reaction forces. Carbon fiber footplates and shank cuffs provided rigid yet comfortable load transfer interfaces. Sensors on the footplate informed a high-level controller used to provide adaptive plantarflexor torque during stance phase and/or constant dorsiflexor torque during swing phase (Fig. 1B). A custom embedded torque transducer at the ankle provided feedback for low-level closed-loop torque control. The total bilateral mass of the device ranged from 2.4 to 2.6 kg, depending on the cable length, and size of the footplates and cuffs (Table 1). The exoskeleton’s peak torque output was 30 Nm. Between 50 and 65% of the total exoskeleton mass (depending on the configuration) was contained within the waist assembly so that the detriment of distally added weight on metabolic power was minimized [22]. Mass minimization, modularity, comfort, and ease of donning and operation were important criteria that guided the design.

Table 1 Exoskeleton mass breakdown

Waist assembly and cable transmission

The waist assembly housed the exoskeleton actuation and control hardware including the motors, custom printed circuit board (PCB), and battery (Fig. 2). A padded harness system fastened the assembly to the waist (Fig. 1A). A modular fiber-reinforced 3D-printed assembly casing was designed to mount two motors (EC4-Pole 90 W with 89:1 GP 22HP gearbox, Maxon) via cartridges and house the electronics module and battery (Fig. 2B). 18 mm sprockets were welded onto the gearbox output shafts and moved chains within the cartridge to actuate the cable transmission that rotated the ankle assembly, transmitting torque and power from the motor to the user. Steel cables were looped through the chain ends and were held in place by a guide and swage (Fig. 2C). Nylon webbing restraints held the cable transmission system to the thighs (Fig. 1A). Each motor mounting cartridge was removable, allowing for quick and easy replacement of each exoskeleton leg assembly independently. The cartridges were designed for ease of maintenance and to house different motor configurations and sizes. They could be quickly swapped for taller or shorter cable configurations depending on the user. Transmission cable configurations were made in set sizes to span a range of user heights (< 160 cm, 160–185 cm,  > 185 cm). The custom PCB interfaced with sensing, control, and wireless communication hardware including a microcontroller (Teensy 3.6, PJRC), motor drivers (ESCON Module 50/8, Maxon), Bluetooth module, and other components to regulate battery voltage and amplify measurement signals (e.g., INA125P, Texas Instruments). A 5 V cooling fan provided airflow through the motor assembly (Fig. 2A). A 24 V, 2000 mAh Li-Ion battery (KamPing) for this study was selected to provide an ambulatory duration equal to or greater than the typical physical therapy session (20–35 min [23, 24]) when walking near the peak torque rating.

Fig. 2
figure 2

Waist assembly overview. A Closed and open pictures of the waist assembly module and harness system. The waist assembly module housed the motors, motor cartridges, PCB, battery, and wiring harness. B Assembled view of a motor cartridge assembly. C Exploded view of a motor cartridge assembly. (1) A 90 W Maxon motor with an 89:1 gearbox. (2) An 8-tooth sprocket welded onto the gearbox output shaft. (3) Reinforced motor blocks were the interface between the motor and cartridge. (4) A chain driven by the sprocket actuated the cable transmission. (5) A sliding cover on each cartridge permitted easy access to the chain assembly (6) A thrust bearing supported the motor shaft to prevent tip deflection during operation. (7) The cartridge was 3D-printed and reinforced with carbon fiber aligned with the long axis. (8) The Bowden sheaths guided steel cables down to the ankle assembly. (9) Wire strain relief. (10) Steel cable looped through final link on each side of the chain and passed through the Bowden sheath. (11) Steel bolts held the motor subassembly in place within the cartridge and attached the cartridge to the rest of the motor assembly. (12) A crimped swage held the steel cable looped through the chain. (13) A small guide component prevented cable stress concentrations and failure

Ankle assembly

The ankle assembly was designed to minimize distal mass and lateral protrusion from the shank, support cable transmission reaction forces, and provide a rigid interface to support and assist a user during activity. Mass added distally on the body increases the metabolic cost of walking more than when it is placed more proximally [22] and limits an exoskeleton’s theoretical potential for benefit [25]. Components placed on the medial portion of the lower limb increase the risk of inter-limb collisions, while posterior or lateral protrusions may cause collisions with the environment. Our previous prototypes suffered from a lack of assembly stiffness due to a large moment arm between the user and the lateral upright and the absence of out-of-plane stiffening geometry [16, 18, 26]. We addressed these issues through mechanical design and material selection specifically intended to maximize assembly stiffness, such as using a square carbon fiber tube for the upright, incorporating stiffening ridges to the footplate and cuff, and reducing lateral protrusion of the ankle assembly by designing a low-profile ankle joint with custom sensors (Fig. 3).

Fig. 3
figure 3

Ankle Assembly. A Assembled view of an entire ankle assembly, including calf cuff, machined carbon fiber upright, tensioners, instrumented exoskeleton joint, and footplate. Tensioners compressed the Bowden sheath via a pull-and-twist knob and kept the cable transmission taut. B Exploded view of the exoskeleton joint. (1) Torque transducer (2) Strain gage. (3) Carbon fiber-reinforced pulley. (4) Steel cable transmission crimping site. (5) Thrust ball bearings. (6) 6 mm shoulder bolt. Four steel bolts fixed the torque sensor to the pulley. (7). Angle sensor assembly and exploded view. A gear shaft meshed with the pulley rotated a diametric magnet underneath a stationary Hall sensor. The resulting voltage was used to calculate joint angle and angular velocity. (8) Removable pulley bridge allowing assembly within the carbon fiber tube. (9) FSR connection

The ankle assembly incorporated a single degree of freedom rotational joint and interfaced with the user via a shank or calf cuff and a footplate (Fig. 3A). The steel cables rotated a torque- and angle-measuring pulley assembly (Fig. 3B). The pulley was placed within a carbon fiber tube and was supported on both ends by flanged bearings (Fig. 3B). The pulley was 80 mm in diameter and formed a 5:1 gear reduction with the motor sprocket. While the pulley design permitted 120 degrees of motion before colliding with the upright, the chain assembly limited the motion to 80 degrees which was sufficient to capture the biological ankle range of motion [27]. We designed a custom low-profile in-line torque transducer for low-level motor control and torque measurement (Fig. 3B) with the goal of minimizing the physical profile of the assembly and lateral lever arm. The lateral lever arm in this design was 3 cm measured from the center axis of the upright to the edge of the footplate vs. 5 cm for a previous prototype [16, 18, 28]. The lateral lever arm, and consequently both the coronal bending and axial twisting moments, were 40% smaller than on our previous devices. We also designed a custom embedded angle sensing unit that resided above the pulley within the carbon fiber tube to provide a platform for the development of new angle- or velocity-dependent control strategies, patient monitoring of ankle angle or range of motion, and measurement of the device’s mechanical power. The footplate was designed to be rigid but lightweight and had a curved feature to match the shape of the foot and metatarsals during toe-off [26]. Cuffs and footplates were made in set sizes, were easily swappable, and each footplate size spanned several shoe sizes [26]. A force-sensitive resistor (FSR, Flexiforce A502, Tekscan) placed on the footplate spanning the 1st through 3rd metatarsal heads under the ball of the foot was used by the microcontroller to detect gait events and generate real-time stance torque profiles.

The following subsections detail specific experiments related to the hardware validation, human performance testing, and usability assessment portions of the experimental protocol summarized in Fig. 1C.

Torque sensor design

Our custom torque transducer was a machined 7075-T651 aluminum part instrumented with strain gages designed to bi-directionally measure up to 30 Nm of torque (Figs. 3B,  4A). The thickness of the transducer was 10 mm and the mass was 30 g. For comparison, commonly used low-profile commercial sensors are over 25 mm thick and weigh over 50 g (e.g., Transducer Techniques TRT-500, [16, 18, 28,29,30]). The transducer measured the sagittal bending moment generated between the cable-driven pulley and footplate (Fig. 3, Additional file 1: Fig. S1). The full Wheatstone bridge strain gage configuration minimized the effects of temperature and out-of-plane loading [31], isolating sagittal-plane torque applied to the user’s ankle joint. The Wheatstone bridge voltages were measured, summed, and amplified using a differential op-amp and a 1-kOhm resistor (INA125P, Texas Instruments) on our custom PCB. Refer to the Torque Sensor Validation section in Additional file 1 for methods and figures related to the experimental setup for validating the torque sensor measurement and assessing its ability to isolate sagittal plane moments.

Fig. 4
figure 4

Torque (A) and angle sensor (BD) validation results. A Linear regression for estimating torque applied to the transducer given a voltage measurement with root-mean-squared error (RMSE). Refer to Additional file 1 for torque sensor sensitivity to out-of-plane loads. B Linear regression relating angle sensor output to motion capture with RMSE; a positive angle corresponds to plantarflexion (PF). C Time series angle measurement with RMSE. D Comparison of sensor-estimated joint velocity to the motion capture result. Refer Additional file 1 for a comparison of sensor and motor velocity estimates during validation (Additional file 1: Fig. S5B)

Angle sensor design

An angle sensor was located above the pulley axis of rotation, residing within the carbon fiber tube to minimize the lateral protrusion of the assembly. The assembly consisted of a 3D-printed plastic gear shaft enclosed in hubs that meshed with a gear on the pulley (Fig. 3B). The gear shaft rotated a diametric magnet underneath a Hall Effect sensor (SS49E, Honeywell) in the lateral hub. A rotating diametric magnet in this configuration induced a repeatable sinusoidal voltage response from the Hall sensor [32]. The angle measurement was then used to estimate joint angular velocity in real time by numerical differentiation (Additional file 1: Fig. S4). Refer to the Angle Sensor Validation section of Additional file 1 for methods and figures related to the experimental setup for validating the sensor angle and velocity measurements.

Software and control

The FSR signal was used to detect transitions between stance and swing, and determine an assistive torque profile during stance. A software threshold defined as a percentage of the total FSR signal range could be increased or decreased to adjust the initiation of stance phase and swing phase assistance. We used a high-level proportional joint moment controller (PJMC [19], Fig. 1) to generate an adaptive torque profile (\(T_{set} \left( t \right))\) proportional to a real-time estimate of the biological ankle moment during stance phase, as in Eq. 1:

$$T_{set} \left( t \right) = T_{0} \frac{{F_{FSR} \left( t \right)}}{{F_{cal} }}$$
(1)

where \(F_{FSR} \left( t \right)\) was the instantaneous FSR reading, \(F_{cal}\) was the reference calibration value defined by the average of the peak FSR reading over three steps for each leg, and \(T_{0}\) was the desired peak exoskeleton torque (e.g., 30 Nm). The footplate FSR captured the shape and magnitude of a signal that served as an estimate of the total biological ankle moment [19, 33]. \(F_{cal}\) normalized the FSR signal, so any variance in the signal magnitude due to FSR placement on the footplate or foot contact with the FSR was eliminated and didn’t affect the real time torque profile \(T_{set} \left( t \right)\). Nominal constant dorsiflexor assistance could be applied during swing phase. A low-level PD controller tracked the generated torque profile using measurements from the torque sensor at the ankle joint (Fig. 1B). PJMC was recently validated across variable terrain including inclined treadmill walking and stair ascent [33], and allows users to seamlessly transition between terrains. The same PJMC parameters were used for all participants and terrains in this study.

Unimpaired cohort experiments

We recruited six healthy adults spanning a range of body sizes with the goal of demonstrating applicability of our device to unimpaired moderate-intensity walking performance augmentation (Table 2). During the experiment, participants walked for six minutes with and without exoskeleton assistance on a treadmill with a five-degree incline. We used a five-degree incline to mimic the maximum allowable ramp angle from Americans with Disabilities Act (ADA) guidelines [34]. We selected moderate intensity incline walking for our unimpaired performance testing experiment primarily because we believe it reflects the real utility of ankle exoskeleton in unimpaired populations, namely, augmenting moderate- to high-intensity ambulatory activities that have elevated ankle plantarflexor demand. Additionally, this condition satisfied our goal of demonstrating a potential benefit beyond the most commonly investigated terrain (level ground). Participants used the shortest exoskeleton configuration that allowed them to walk without limiting step length. Participants were also sized for footplates and cuffs that fit snugly and were comfortable. Proper footplate fit was qualified by contact between the ball of the foot and the footplate FSR and by close alignment (within 3 cm) of exoskeleton and biological ankle joint centers. Footplate mounting hole patterns allowed for easy joint center alignment. Extra foam padding was added to footplates and cuffs as needed for comfort and fit.

Table 2 Unimpaired participant information
Fig. 5
figure 5

Representative exoskeleton measured (blue) and prescribed (red) torque (top row), velocity (middle row), and power (bottom row) profiles (mean ± standard deviation) from a single subject (P1, Table 2) during assisted walking on a 5-degree incline (left column) and on a stair-stepping machine (right column). Mechanical power (bottom row) was calculated by multiplying measured exoskeleton torque (Nm) and angular velocity (radians per second). Torque and power were normalized by the participant’s body mass. Torque tracking error during early stance and immediately after toe-off are due to motor torque rate and speed limitations. Refer to Additional file 2 for additional participant torque, velocity, and power curves and to Additional file 3 for a short section on power consumption and electrical to mechanical power efficiency

Prior to the first trial, participants were given 10–15 min of exoskeleton acclimation time during which an operator tuned the exoskeleton assistance and treadmill speed. After a standing torque sensor zero calibration, participants walked at 1.0 m/s with 0.35 Nm/kg of nominal peak stance phase assistance and 0.05 Nm/kg of swing phase assistance while the dynamic FSR calibration was performed. Torque levels were chosen from our prior works [18, 21, 28] and pilot tests. Then, the operator increased the treadmill speed until the participant confirmed that the activity was of moderate intensity (Table 2). The operator re-calibrated \(F_{cal}\) to ensure good torque tracking at the faster walking speed, modified the FSR state transition threshold to ensure timely transitions between stance and swing (if needed), and adjusted swing phase assistance until the participant confirmed that the dorsiflexor torque was helpful after toe-off but did not impede the following heel strike. Most participants were comfortable with exoskeleton assistance after about 5–10 min of acclimation time. The calibrated exoskeleton parameters were saved and used during the shod-exoskeleton comparison experiment.

Each participant was assigned one shod and one assisted trial; trial order was alternated across the cohort (Table 2). We collected metabolic data using an indirect calorimetry unit (K5, COSMED). Oxygen and carbon dioxide volumes were used to calculate metabolic power using Brockway’s equation [35] for the last 3 min of each trial [36, 37]. Prior to each trial, the participant stood quietly for 2–3 min or until respiratory data were steady. The last minute of respiratory data during standing prior to each trial was used to calculate basal metabolic rate. The metabolic power for each walking trial was offset by the basal rate and normalized by body mass to calculate net metabolic power [38]. Between trials, participants sat and rested for 10 min.

We streamed exoskeleton signals, including motor current and velocity, desired and actual joint torques, and exoskeleton joint angle and angular velocity, to a custom MATLAB (R2018b, MathWorks) interface at 100 Hz. Exoskeleton joint power was calculated as the product of the measured joint torque and angular velocity for each leg. The net metabolic powers between the shod and exoskeleton trials were compared to assess the impact of powered exoskeleton assistance on energetics.

Impaired cohort experiments

We recruited seven individuals with CP spanning a range of ages and impairment levels with the goal of demonstrating that our device can be effective at improving aerobic capacity during a maximal exertion test in this patient population (Table 3). Moving beyond our prior research that focused on augmenting walking on level ground in CP, we sought to explore application of ankle exoskeleton assistance to improve maximum exertion performance, which has not been previously explored in the literature. We designed an experiment to test performance on a stair-climbing machine as a way to expanding our understanding on the use of ankle assistance across different terrains and ambulatory intensities. Individuals with CP have difficulty with stair ascent [39] and are acutely susceptible to lower leg muscle fatigue [40], so we sought to demonstrate that our device was effective at prolonging the duration of this high-intensity activity. Inclusion criteria for this experiment included diagnoses of CP; the ability to walk on a stair machine for at least 5 min; Gross Motor Function Classification System (GMFCS) level I, II, or III; at least 20° of passive ankle plantarflexion range of motion; no knee extension or ankle dorsiflexion contractures greater than 15°; no orthopedic surgery completed in the prior 6-month period; and the absence of any medical condition other than CP that would affect safe participation. Participants were fitted with an ankle exoskeleton, footplates, and cuffs as described in the previous section. The same exoskeleton calibration procedure as in the previous section was conducted prior to exoskeleton-assisted trials.

Table 3 Impaired participant information

The maximal exertion test protocol was as follows. The stair stepping rate was increased from each participant’s comfortable rate by one intensity level (0.3–0.4 floors/min) every thirty seconds until the participant indicated they wanted to stop. All participants wore a safety harness and were surrounded by researchers ready to stop the machine and support the participant to prevent harm. Participants completed one shod and one exoskeleton-assisted maximal exertion trial (Additional file 4: Video S1). Participants took a 20-min break between trials and confirmed that they were fully rested. Two of the seven participants were too light to trigger an increase in stair stepping rate and were unable to complete the experiment. We prescribed 0.30 Nm/kg of nominal peak plantarflexor assistance and 0.03 Nm/kg dorsiflexor assistance. During each trial, we recorded duration, step rate, and metabolic rate. We calculated the total distance travelled in number of floors (1 floor = 16 steps). After each trial, we recorded each participant’s perceived exertion using standard scales [41]. We compared the floors ascended between the shod and exoskeleton conditions. We also compared net metabolic power between the conditions over the duration of the shortest trial because intensity increased as the trials continued and we sought to make a direct comparison of metabolic power for the same duration and intensity. For example, if the shod trial was 5 min long and the assisted trial was 7 min long, we compared the average net metabolic power across the first 5 min of both trials.

Usability assessment

All seven participants with CP performed a device usability assessment with the goal of demonstrating that time to don and operate the device improved with practice and could be completed in less than 5 min. Our usability experiments were motivated by our interest in conducting future evaluations of ankle exoskeleton assistance for augmenting mobility in free-living scenarios. The cohort included a wide range of participants as we were interested in receiving a variety of feedback from both children and adults on the usability and effectiveness of our exoskeleton. We recorded the time of each step of the donning process and the total app setup time (from powering on the device to walking with full torque magnitude) for each participant, including time spent reading instructions. Participants completed the exercise three times. Three individuals from the unimpaired cohort (P1, P3, and P4) were selected to perform the same assessment as a reference. Donning our device was a three-step process. Participants (1) prepared the device by placing it on a chair and placing footplates within shoes, (2) inserted their feet into the shoes, tied the shoes tightly, and strapped the cuffs in place on each shank, and (3) clipped and adjusted the waist assembly straps (Additional file 4: Video S2). We designed a custom iOS application that would automatically connect to the exoskeleton and guide the user through steps needed to start walking with assistance. Controlling the device using the iOS app was a quick three-step process that included: (1) user weight input in pounds, (2) static torque transducer calibration during quiet standing, and (3) dynamic controller calibration (state detection and ankle moment normalization) while walking in zero-torque mode. When ready, participants were instructed to walk and the firmware automatically completed the walking calibration and provided 0.30 Nm/kg of nominal peak stance torque and 0.03 Nm/kg swing torque, building in magnitude from zero over the course of three steps per leg. Refer to Additional file 4 for a link to the donning instructions on our website. The most affected participant (CP4) received parental assistance due to severe upper-extremity disability. To assess subjective user experience and quantify the usability of the exoskeleton, each participant completed the System Usability Scale questionnaire [42]. Briefly, the System Usability Scale includes 10 statements rated by means of a 5-point Likert scale, from 1 (strongly disagree) to 5 (strongly agree), and the scores have a range of 0 to 100 that is divided into five scales: score of 0–25: worst, score of 25–39: poor, score of 39–52: OK, score of 52–85: excellent, and score of 85–100: best imaginable [43].

Statistics

All data sets were tested for normality using Shapiro–Wilk tests at the 5% significance level [44] and all samples were normally distributed. We compared net metabolic power for the unimpaired cohort and net metabolic cost of transport and distance travelled for the impaired cohort between shod and exo-assisted trials. For the usability section, we compared time needed to don and setup the device across three attempts. Two-tailed paired t-tests were used to assess differences at 5% significance for all group-level comparisons. Cohen’s d (d) was used to calculate effect size as the difference of group means divided by the pooled standard deviation, where 0.2 was considered a small effect, 0.5 a medium effect, and 0.8 a large effect [45]. All analyses and statistical comparisons were done in MATLAB. Simple statistical comparisons were used without p-value corrections so that the reader may judge the significance and impact of the group-level comparisons for themselves.

Results

Torque sensor validation

The linear model relating our custom torque sensor’s voltage output to applied torque explained 99% of the data variance and had low overall mean testing error and variance between predicted and actual torques (Fig. 4A). The torque absolute test root-mean-squared error (RMSE) was 0.65 Nm. We assessed the ability of the torque sensor to isolate sagittal plane moments and confirmed that out-of-plane sensitivity was between 5.2% and 15.5% of the sagittal bending moment sensitivity (Additional file 1: Fig. S3). No evidence of fatigue or offset drift has been observed.

Angle sensor validation

The linear model relating our custom angle sensor’s voltage output to the motion capture angle explained 99% of the data variance and had low overall mean testing error and variance between predicted and actual pulley angles (Fig. 4B). The angle and angular velocity absolute RMSE computed from the ankle joint sensor relative to motion capture was 0.67 degrees and 9.01 deg/s, respectively, for a 2.5-s sample (Fig. 4CD). Estimating the exoskeleton joint velocity using measured motor velocity and the transmission system gear reductions yielded large error compared to motion capture (RMSE = 17.85 deg/s, Additional file 1: Fig. S4B).

Unimpaired cohort experiments

Five of the six unimpaired participants responded well to exoskeleton assistance during inclined treadmill walking (Fig. 6). Improvements in energy cost during the last 3 min of the trial ranged from 7.4% to 18%, with one participant (P5) showing no change when walking with assistance. Our cohort had a 9.9 ± 2.6% (mean ± standard error) improvement in metabolic power when walking with vs without the device (p = 0.012, d = 0.59, Fig. 6). Exoskeleton torque, angular velocity, and mechanical power were captured for five of the six unimpaired participants (Fig. 5, Additional file 2).

Fig. 6
figure 6

Unimpaired cohort metabolic results during incline treadmill walking with and without the exoskeleton. Five of the six participants responded well to exoskeleton assistance and showed a reduction in net metabolic power compared to no device. Refer to Table 2 for participant information and trial order. *Significant at 95% confidence

Impaired cohort experiments

All participants were able to safely complete the maximal exertion stair-climbing test without incident. The number of floors climbed during the maximal exertion experiment increased by 38.4 ± 23.6% with exoskeleton assistance compared to shod (p = 0.013, d = 0.25, Table 4, Fig. 7A). Despite the increase in distance ambulated, the net metabolic powers for assisted and unassisted maximal exertion tests were not significantly different (p = 0.49, Fig. 7B), and perceived exertion on a scale of 1–10 was similar (6.8 ± 0.8 with the device vs 7.2 ± 1.5 without the device, p = 0.50).

Table 4 Maximum exertion speed and distance results
Fig. 7
figure 7

Maximal exertion stair-climbing results (top row) and usability assessment (bottom row). A Number of floors climbed during the maximal exertion experiment. B Average net metabolic power during the max exertion experiment. In general, participants had little change in energy cost but climbed more floors when walking with the device. 1CP1 had a large reduction in net metabolic power without a change in floors travelled suggesting that he likely could have climbed higher. C Device don time. The final attempt time was significantly lower than both the first and second attempts. 2A caregiver assisted CP4 throughout the assessment. 3CP5 had spastic hemiplegia of the upper and lower-extremities and had difficulty handling the device, but had a nearly three-minute improvement with practice. D App setup time. Setup time decreased with practice to ~ 30 s. E App at each step. From left to right: (1) start screen checks to see if a device is in range, (2) optional donning instructions (link available in Additional file 4), (3) mass input, (4) torque sensor calibration, (5) screen for adjusting parameters and starting trial, (6) example of an active trial

Usability assessment

The usability assessment results showed that total don time and app setup time generally decreased with practice (Fig. 7DE). Placing the footplates into the shoe and donning the ankle assembly was the most time- consuming step. The third attempt time (3.5 ± 0.7 min) was significantly different from the first attempt (down 100 ± 30 s, p = 0.018, d = 0.66) and from the second attempt (down 58 ± 14 s, p = 0.006, d = 0.51). Refer to Additional file 4: Video S2 for an example of the usability assessment. For reference, the average unimpaired final don time was 1.5 ± 0.2 min. The final app setup time (28 ± 6 s) was significantly different from the second attempt (down 24 ± 8 s, p = 0.016, d = 1.26). For reference, the average unimpaired final app setup time was 14 ± 1 s. The average System Usability Scale questionnaire score of the impaired cohort was 81.8 ± 8.4 (“excellent”).

Discussion

The goal of this study was to design a highly useable lightweight ankle exoskeleton with custom sensing and provide an initial indication of its effectiveness across a range of terrains and users. Our design utilized a modular motor assembly mounted at the waist to provide ankle torque via interchangeable Bowden cables, calf cuffs, and footplates in standard sizes that spanned a wide range of users from children to adults. We validated our in-line sensing modules that provided direct, real-time measurement of ankle angle, torque and power; these sensors provide an opportunity for the development of new control strategies (e.g., velocity-dependent muscle force prediction), patient monitoring (e.g., ankle range of motion), closed-loop torque control, and assessment of mechanical performance (e.g., torque tracking). The device was able to provide up to 30 Nm of peak torque with a mass 2.3–2.6 kg depending on size and battery selection. We developed a custom iOS application allowing users to control the device themselves. We accept our primary hypotheses that (1) improvements in ambulatory performance (economy or distance) for both unimpaired and impaired cohorts are detectable during diverse exoskeleton-assisted tasks, and (2) users can don and control the device in less than 5 min. While the improvements in walking economy and distance are motivating, we caution the reader to interpret the results with care due to the small cohort sizes.

Our battery-powered device improved moderate-intensity incline walking efficiency by 10% compared to without wearing the device in an unimpaired cohort of adults of diverse body masses and statures (Fig. 6). Compared to untethered and even tethered systems, our device had competitive peak torque capability when normalized to the mass added onto the shank (Additional file 2: Fig. S2) and demonstrated similar improvements in energy expenditure. For example, the seminal untethered, shank-mounted exoskeleton from Mooney et al., weighed 1.06 kg per leg, produced up to 45 Nm (0.5 Nm/kg) peak torque during unloaded, level walking for a small cohort of heavy unimpaired adults (n = 6, mass = 89 ± 8 kg, mean ± SD) and induced a group-level reduction in energy expenditure of 11 ± 4% (mean ± SE) compared to walking without wearing the device [3]. On the other side of the spectrum, an untethered soft exosuit from Awad et al. with a peak torque of about 15 Nm (0.15 Nm/kg) and a distal mass of 0.42–0.50 kg was successful in preliminary and clinical trials during unilateral assistance with stroke survivors [1, 17]; we are not aware of bilateral unimpaired metabolic results for this system. A tethered ankle exoskeleton from Galle et al. had a peak torque of 36.6 Nm (0.6 Nm/kg), a distal mass of 0.445 kg per leg, and induced a group-level reduction of 12.3 ± 2.9% compared to shod (n = 14, mass = 61.0 ± 4.5 kg) [14]. A high-powered tethered exoskeleton from Zhang et al. with distal mass 0.875 kg induced a 21% improvement in a single subject during walking at a 10-degree incline [15] whereas, in the present study, the maximum observed reduction in energy cost during 5-degree incline walking was 18%. Compared to the aforementioned exoskeletons, our device produced up to 30 Nm, had a distal mass of 0.415 kg per leg, and induced a similar group-level reduction in energy consumption in an unimpaired cohort with greater mass variability than the aforementioned studies (mass range of 46–96 kg). As far as we are aware, our device had one of the lightest ankle assemblies of any research or commercial powered device at the time of this writing. The light distal mass likely contributed significantly to the observed benefits in energy reduction based on the augmentation factor, an estimate of potential benefit that balances positive power production with detriments to energy consumption due to added mass [25, 46]. Though our exoskeleton provided less torque and power than the device described in Mooney et al. [3], it had a similar augmentation factor due to its reduced distal mass (between 28 and 56 W vs. 44 W in [46] and 33 W in [25]). Augmentation factor was calculated as described in Mooney et al. [25] using our in-line joint sensors. Refer to Additional file 5 for an example calculation of augmentation factor.

Maximal exertion tests are commonly used to assess functional capacity, with individuals with CP having lower maximal oxygen consumption compared to unimpaired individuals [47]. In this study, we demonstrated that children and young adults with CP were able to ascend almost 40% more steps on a stair stepping machine while using the device during a maximal exertion test. This provides new insight into the potential for wearable ankle assistance to provide both psychological and physiological benefits during high-intensity activities (Fig. 7CD). For example, average metabolic power or perceived exertion remained the same despite the encouraging group-level increase in distance. One of the most promising findings was that our most impaired participant (CP4) ascended nearly twice as far with exoskeleton assistance compared to no device (2.1 vs. 0.9 floors, respectively). An interesting result at the individual level was that while CP1 had only a very small improvement in distance when using the device, there was a sizeable reduction in metabolic cost compared to no device (~ 23%, Fig. 7B), suggesting that he opted to decrease his effort while using the device as opposed to maintaining the same level of effort when not using the device like most participants. While additional research is needed, our results suggest that lightweight untethered ankle assistance may allow for increased and task-oriented training in individuals with CP, which has been proven successful at improving mobility in both CP and post-stroke patients [48,49,50].

To the best of our knowledge, this is the first study to report on the design of an untethered exoskeleton that allows individuals with physical disabilities to don and operate the device without researcher intervention. Individuals with CP were able to don and control the device to the point of walking with assistance (either on their own or with caregiver aid) in an average of just 4 min, suggesting that future deployment in clinical and home settings may be realistic. While the mobile app allows the user to modify exoskeleton tuning parameters, we envision that the user would not need to adjust tuning parameters after the initial device fitting and tuning process with a trained physical therapist or researcher. The rate of improvement over three attempts suggests that donning and setup time would likely decrease further with continued use. For example, one participant (CP5), a determined teen with upper- and lower-extremity spastic hemiplegia, had considerable difficulty donning the device on his own but improved his time by almost 3 min with practice (Fig. 7D). When prompted for feedback on the design, comfort, and control of the exoskeleton, users and caregivers tended to comment favorably on the low hardware profile on the shank and waist, low device weight, shank and waist interface comfort, quick don and doff times, and intuitive torque assistance. Donning the shoes and cuffs was similar to typical AFOs and the waist straps only needed to be adjusted once. The System Usability Scale questionnaire score of 81.8 ± 8.4 (excellent) was comparable to a commercial robotic device (Ekso GT, score: 83) and better than another pre-market devices (PASfinal, score: 59.5) [51].

While small cohort exoskeleton studies are common [3, 13, 14, 25, 52], we acknowledge that a limitation to this study was the small sample sizes used for statistical analysis and encourage the reader to interpret group-level results and statistical significance with caution. Another limitation of this study was that the battery used during the experiments provided just 37 min of walking duration at the average torque used across both cohorts (~ 22 Nm). While we met and exceeded our goal of completing performance testing with a battery meeting the typical ambulatory duration of a standard physical therapy session (20–35 min [23, 24]), the battery used in this study would be insufficient for all day use. However, we theorize that in the future deployment of this or similar commercial exoskeleton systems, users may prefer to minimize adding mass to the body by interchanging multiple batteries throughout the day.

Conclusion

This study reported on the design and initial testing of a new, lightweight ankle exoskeleton. We validated custom low-profile joint-level sensing and achieved one of (if not) the lightest reported distal mass placements of any tethered or untethered powered ankle exoskeleton. The device performed well across a range of ambulatory conditions, walking abilities, and body sizes, reducing incline walking energy consumption in unimpaired adults and improving maximal exertion stair climbing distance in children and adults with CP. Our device was effective for children with CP as light as 38 kg and unimpaired adults as heavy as 92 kg, suggesting that 0.30–0.35 Nm/kg peak plantarflexor assistance and 0.03–0.05 Nm/kg dorsiflexor assistance was sufficient to improve ambulatory performance. All of our unimpaired participants and most of our participants with CP were able to don and initiate user-calibrated ankle assistance in under 5 min without researcher intervention. Future work will continue to explore the effectiveness of adaptive ankle exoskeleton assistance across a multitude of challenging terrains for unimpaired and impaired participants.

Availability of data and materials

The datasets used and analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

CP:

Cerebral palsy

PCB:

Printed circuit board

GMFCS:

Gross Motor Function Classification System

RMSE:

Root-mean-squared error

PF:

Plantarflexion (refers to angle)

DF:

Dorsiflexion (refers to angle)

References

  1. Awad LN, Bae J, O’Donnell K, De Rossi SMM, Hendron K, Sloot LH, et al. A soft robotic exosuit improves walking in patients after stroke. Sci Transl Med. 2017;9(400):1–12.

    Article  Google Scholar 

  2. Galle S, Derave W, Bossuyt F, Calders P, Malcolm P, De Clercq D. Exoskeleton plantarflexion assistance for elderly. Gait Posture. 2017;52:183–8.

    Article  CAS  Google Scholar 

  3. Mooney LM, Herr HM. Biomechanical walking mechanisms underlying the metabolic reduction caused by an autonomous exoskeleton. J Neuroeng Rehabil. 2016. https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-016-0111-3.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Witte KA, Fiers P, Sheets-Singer AL, Collins SH. Improving the energy economy of human running with powered and unpowered ankle exoskeleton assistance. Sci Robot. 2020. https://0-doi-org.brum.beds.ac.uk/10.1126/scirobotics.aay9108.

    Article  PubMed  Google Scholar 

  5. Neptune RR, Sasaki K, Kautz SA. The effect of walking speed on muscle function and mechanical energetics. Gait Posture. 2008;28(1):135–43.

    Article  Google Scholar 

  6. Farris DJ, Sawicki GS. The mechanics and energetics of human walking and running: a joint level perspective. J R Soc Interface. 2012;9(66):110–8.

    Article  Google Scholar 

  7. Franks PW, Bryan GM, Martin RM, Reyes R, Collins SH. Comparing optimized exoskeleton assistance of the hip, knee, and ankle in single and multi-joint configurations. BioRxiv. 2021. https://0-doi-org.brum.beds.ac.uk/10.1101/2021.02.19.431882.

    Article  Google Scholar 

  8. Lin PY, Yang YR, Cheng SJ, Wang RY. The relation between ankle impairments and gait velocity and symmetry in people with stroke. Arch Phys Med Rehabil. 2006;87(4):562–8.

    Article  Google Scholar 

  9. Davids JR. The foot and ankle in cerebral palsy, vol 41. Orthopedic Clinics of North America. 2010. p. 579–93.

  10. Barber L, Barrett R, Lichtwark G. Medial gastrocnemius muscle fascicle active torque-length and Achilles tendon properties in young adults with spastic cerebral palsy. J Biomech. 2012;45(15):2526–30.

    Article  Google Scholar 

  11. Dallmeijer AJ, Rameckers EA, Houdijk H, de Groot S, Scholtes VA, Becher JG. Isometric muscle strength and mobility capacity in children with cerebral palsy. Disabil Rehabil. 2017;39(2):135–42.

    Article  Google Scholar 

  12. Yılmaz Topçuoğlu MS, Krautwurst BK, Klotz M, Dreher T, Wolf SI. How do children with bilateral spastic cerebral palsy manage walking on inclines? Gait Posture. 2018;66:172–80.

    Article  Google Scholar 

  13. Malcolm P, Derave W, Galle S, De Clercq D. A simple exoskeleton that assists plantarflexion can reduce the metabolic cost of human walking. PLoS ONE. 2013;8(2):e56137.

    Article  CAS  Google Scholar 

  14. Galle S, Malcolm P, Collins SH, De Clercq D. Reducing the metabolic cost of walking with an ankle exoskeleton: interaction between actuation timing and power. J Neuroeng Rehabil. 2017. https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-017-0235-0.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Zhang J, Fiers P, Witte KA, Jackson RW, Poggensee KL, Atkeson CG, et al. Human-in-the-loop optimization of exoskeleton assistance during walking. Science. 2017;356(6344):1280–3.

    Article  CAS  Google Scholar 

  16. Lerner ZF, Harvey TA, Lawson JL. A battery-powered ankle exoskeleton improves gait mechanics in a feasibility study of individuals with cerebral palsy. Ann Biomed Eng. 2019;47(6):1345–56.

    Article  Google Scholar 

  17. Awad LN, Esquenazi A, Francisco GE, Nolan KJ, Jayaraman A. The ReWalk ReStoreTM soft robotic exosuit: a multi-site clinical trial of the safety, reliability, and feasibility of exosuit-augmented post-stroke gait rehabilitation. J Neuroeng Rehabil. 2020. https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-020-00702-5.

    Article  PubMed  PubMed Central  Google Scholar 

  18. Orekhov G, Fang Y, Luque J, Lerner ZF. Ankle exoskeleton assistance can improve over-ground walking economy in individuals with cerebral palsy. IEEE Trans Neural Syst Rehabil Eng. 2020;28(2):461–7.

    Article  Google Scholar 

  19. Gasparri GM, Luque J, Lerner ZF. Proportional joint-moment control for instantaneously adaptive ankle exoskeleton assistance. IEEE Trans Neural Syst Rehabil Eng. 2019;27(4):751–9.

    Article  Google Scholar 

  20. Gasparri GM, Bair MO, Libby RP, Lerner ZF. Verification of a robotic ankle exoskeleton control scheme for gait assistance in individuals with cerebral palsy. In: IEEE International Conference on Intelligent Robots and Systems; 2018. p. 4673–8.

  21. Fang Y, Orekhov G, Lerner ZF. Adaptive ankle exoskeleton gait training demonstrates acute neuromuscular and spatiotemporal benefits for individuals with cerebral palsy: a pilot study. Gait Posture; 2020.

  22. Browning RC, Modica JR, Kram R, Goswami A. The effects of adding mass to the legs on the energetics and biomechanics of walking. Med Sci Sports Exerc. 2007;39(3):515–25.

    Article  Google Scholar 

  23. Begnoche DM, Pitetti KH. Effects of traditional treatment and partial body weight treadmill training on the motor skills of children with spastic cerebral palsy: a pilot study. Pediatr Phys Ther. 2007;19(1):11–9.

    Article  Google Scholar 

  24. Kuys SS, Ada L, Paratz J, Brauer SG. Steps, duration and intensity of usual walking practice during subacute rehabilitation after stroke: an observational study. Brazilian J Phys Ther. 2019;23(1):56–61.

    Article  Google Scholar 

  25. Mooney LM, Rouse EJ, Herr HM. Autonomous exoskeleton reduces metabolic cost of human walking during load carriage. J Neuroeng Rehabil. 2014. https://0-doi-org.brum.beds.ac.uk/10.1186/1743-0003-11-80.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Babers J. Optimizing the Mechanical Design of an Untethered Ankle Exoskeleton. Northern Arizona University; 2020.

  27. Hemmerich A, Brown H, Smith S, Marthandam SSK, Wyss UP. Hip, knee, and ankle kinematics of high range of motion activities of daily living. J Orthop Res. 2006;24(4):770–81.

    Article  CAS  Google Scholar 

  28. Lerner ZF, Gasparri GM, Bair MO, Lawson JL, Luque J, Harvey TA, et al. An untethered ankle exoskeleton improves walking economy in a pilot study of individuals with cerebral palsy. IEEE Trans Neural Syst Rehabil Eng. 2018;26(10):1985–93.

    Article  Google Scholar 

  29. Lerner ZF, Damiano DL, Bulea TC. A robotic exoskeleton to treat crouch gait from cerebral palsy: initial kinematic and neuromuscular evaluation. In: Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS. 2016.

  30. Lerner ZF, Damiano DL, Bulea TC. A lower-extremity exoskeleton improves knee extension in children with crouch gait from cerebral palsy. Sci Transl Med. 2017. https://0-doi-org.brum.beds.ac.uk/10.1126/scitranslmed.aam9145.

    Article  PubMed  Google Scholar 

  31. Hoffmann K. Applying the wheatstone bridge circuit. HBM S1569–11 en, HBM, Darmstadt, Ger. 2001.

  32. Honeywell. Hall Effect Sensing and Application. https://www.google.com/url?sa=t&rct=j&q=&esrc=s&source=web&cd=&cad=rja&uact=8&ved=2ahUKEwiS9LSSj6LxAhWM4J4KHb7qBi0QFjABegQIBRAF&url=https%3A%2F%2Fsensing.honeywell.com%2Fhoneywell-sensing-sensors-magnetoresistive-hall-effect-applications-005715-2-en2.pdf&.

  33. Bishe SSPA, Nguyen T, Fang Y, Lerner ZF. Adaptive ankle exoskeleton control: validation across diverse walking conditions. IEEE Trans Med Robot Bionics. 2021;3(3):801–12.

    Article  Google Scholar 

  34. Section 405.2. In: 2010 ADA standards for accessible design. https://www.access-board.gov/ada/guides/chapter-4-ramps-and-curb-ramps/

  35. Brockway JM. Derivation of formulae used to calculate energy expenditure in man. Hum Nutr Clin Nutr. 1987;41(6):463–71.

    CAS  PubMed  Google Scholar 

  36. Sawicki GS, Ferris DP. Mechanics and energetics of level walking with powered ankle exoskeletons. J Exp Biol. 2008;211(9):1402–13.

    Article  Google Scholar 

  37. Sawicki GS, Ferris DP. Mechanics and energetics of incline walking with robotic ankle exoskeletons. J Exp Biol. 2009;212(1):32–41.

    Article  Google Scholar 

  38. MacLean MK, Ferris DP. Energetics of walking with a robotic knee exoskeleton. J Appl Biomech. 2019;35:320.

    Article  Google Scholar 

  39. Sienko Thomas S, Buckon CE, Jakobson-Huston S, Sussman MD, Aiona MD. Stair locomotion in children with spastic hemiplegia: the impact of three different ankle foot orthosis (AFOs) configurations. Gait Posture. 2002;16(2):180–7.

    Article  Google Scholar 

  40. Eken MM, Brændvik SM, Bardal EM, Houdijk H, Dallmeijer AJ, Roeleveld K. Lower limb muscle fatigue during walking in children with cerebral palsy. Dev Med Child Neurol. 2019;61(2):212–8.

    Article  Google Scholar 

  41. Yamauchi SSM. Rating of perceived exertion for quantification of the intensity of resistance exercise. Int J Phys Med Rehabil. 2013. https://0-doi-org.brum.beds.ac.uk/10.4172/2329-9096.1000172.

    Article  Google Scholar 

  42. Brooke J. SUS—a quick and dirty usability scale Usability and context. Usability Eval Ind. 1996;189(194):4–7.

    Google Scholar 

  43. Bangor A, Kortum P, Miller J. Determining what individual SUS scores mean: adding an adjective rating scale. J Usability Stud. 2009;4(3):114–23.

    Google Scholar 

  44. Ghasemi A, Zahediasl S. Normality tests for statistical analysis: a guide for non-statisticians. Int J Endocrinol Metab. 2012;10(2):486–9.

    Article  Google Scholar 

  45. Hillside NJ. Statistical power analysis for the behavioural sciences. Google Scholar. In: Statistical power anaylsis for the behavioral sciences; 1988.

  46. Mooney LM, Rouse EJ, Herr HM. Autonomous exoskeleton reduces metabolic cost of human walking. J Neuroeng Rehabil. 2014. https://0-doi-org.brum.beds.ac.uk/10.1186/1743-0003-11-151.

    Article  PubMed  PubMed Central  Google Scholar 

  47. Unnithan VB, Clifford C, Bar-Or O. Evaluation by exercise testing of the child with cerebral palsy. Sport Med. 1998;26(4):239–51.

    Article  CAS  Google Scholar 

  48. Salem Y, Godwin EM. Effects of task-oriented training on mobility function in children with cerebral palsy. NeuroRehabilitation. 2009;24(4):307–13.

    Article  Google Scholar 

  49. Salbach NM, Mayo NE, Wood-Dauphinee S, Hanley JA, Richards CL, Côté R. A task-orientated intervention enhances walking distance and speed in the first year post stroke: a randomized controlled trial. Clin Rehabil. 2004;18(5):509–19.

    Article  CAS  Google Scholar 

  50. Ma HI, Trombly CA, Robinson-Podolski C. The effect of context on skill acquisition and transfer. Am J Occup Ther. 1999;53(2):138–44.

    Article  CAS  Google Scholar 

  51. Meyer JT, Schrade SO, Lambercy O, Gassert R. User-centered design and evaluation of physical interfaces for an exoskeleton for paraplegic users. In: IEEE International Conference on Rehabilitation Robotics; 2019. p. 1159–66.

  52. Mooney LM, Rouse EJ, Herr HM. Autonomous exoskeleton reduces metabolic cost of walking. In: 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2014; 2014. p. 3065–8

Download references

Acknowledgements

This work was supported in part by the Eunice Kennedy Shriver National Institute of Child Health and Human Development of the National Institutes of Health under award numbers R15HD099664 and R44HD104328. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health. We thank James Babers and Leah Liebelt for their design and manufacturing contributions. We also thank Jeff Yows and Samuel Maxwell for their help with exoskeleton assembly and experiments. We also thank Ben Conner for his help with participant recruitment and experiments.

Funding

This work was supported in part by the Eunice Kennedy Shriver National Institute of Child Health and Human Development of the National Institutes of Health under award numbers 1R01HD107277 and 1R44HD104328. The content is solely the responsibility of the authors and does not necessarily represent the official views of the National Institutes of Health.

Author information

Authors and Affiliations

Authors

Contributions

GO designed the exoskeleton device, led the mechanical and performance validation experiments, analyzed and interpreted data, generated figures, and wrote the manuscript. YF designed and led the maximal exertion and usability experiments, analyzed data, and generated figures. CFC designed the iOS app for controlling the exoskeleton. ZFL was the principal investigator, designed the exoskeleton device and iOS app, conceived, designed, and led the study, interpreted data, and wrote the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Zachary F. Lerner.

Ethics declarations

Ethics approval and consent to participate

All human subjects experiments were approved by Northern Arizona University’s Institutional Review Board under protocol #986744-27 as a part of NCT04119063. All participants 18 and over read and signed informed consent forms. For all minors, we obtained assent and informed written consent from a parent.

Consent for publication

All participants 18 and over read and signed photo and video release forms. For all minors, we obtained assent and written consent from a parent.

Competing interests

ZFL is a co-founder with shareholder interest of a university start-up company seeking to commercialize the device used in this study. ZFL and GO hold intellectual property inventorship rights.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1:

Supplemental figures detailing hardware validation methods and results.

Additional file 2:

Supplemental exoskeleton performance data. P4's torque, velocity, and power curves for inclined walking and stair ascent. Comparison of our device to other prominent exoskeleton studies.

Additional file 3:

Exoskeleton power consumption and electrcal-to-mechanical power efficiency analysis.

Additional file 4:

Video and instruction links. Links to videos of exoskeleton maximal exertion and usability experiments. Link to exoskeleton donning instructions.

Additional file 5:

Augmentation factor calculation for unimpaired cohort.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Orekhov, G., Fang, Y., Cuddeback, C.F. et al. Usability and performance validation of an ultra-lightweight and versatile untethered robotic ankle exoskeleton. J NeuroEngineering Rehabil 18, 163 (2021). https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-021-00954-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-021-00954-9

Keywords