Symmetry function in gait pattern analysis in patients after unilateral transfemoral amputation using a mechanical or microprocessor prosthetic knee

Abstract

Background

Above-knee amputations (AKAs) contribute to gait asymmetry. The level of asymmetry is affected by the type of knee prosthetic module. There is limited evidence suggesting that more technically advanced solutions (microprocessor modules; MicPK) are superior to less advanced ones (mechanical modules; MechPK). The study aimed to evaluate the variable range of hip and pelvic joint movements during gait and look for differentiating areas with an increased level of asymmetry of the gait cycle in individuals who underwent an AKA and are equipped with MicPK or MechPK.

Methods

Twenty-eight individuals participated in the study; 14 were assigned to a study group of individuals who underwent a unilateral AKA, and the other 14 were healthy participants as a control group. The movement task was recorded using the optoelectronic SMART-E system following the standard Davis protocol (the Newington model). A new method of quantifying gait symmetry using the symmetry function (SF) is proposed. SF is an integral measure of absolute differences in time-standardized signals between sides throughout the whole cycle of motion variability.

Results

In the frontal plane, there were significant differences between groups in the asymmetry of the range of movement in the hip joint of the intact limb. In the middle of the support phase, the intact limb was more adducted in individuals with MicPK and less abducted in people with MechPK (differences in mean 180%, p < 0.000; max 63%, p < 0.000; min 65%, p < 0.000). In the sagittal plane, the range of asymmetry of the flexion and thigh extension of the intact limb was similar to and only slightly different from the physiological gait. In the transverse plane, higher asymmetry values were noted for individuals with MicPK. In the final stage of the swing phase, the thigh was more rotated both externally and internally. The size of the asymmetry, when compared to gait of healthy individuals, reached 50% (differences in mean 115%, p < 0.232; max 62% p < 0.26; min 50, p < 0.154).

Conclusions

In the study group, the assessed ranges of pelvic and thigh movement in the hip joint differed only in the frontal plane. Individuals who underwent a unilateral above-knee amputation may gain less from using MicPK than anticipated.

Introduction

The percentage of above-knee amputations (AKAs) performed for reasons other than vascular has remained constant over the years and reached 0.92 amputations per 1000 people [1, 2]. A significant number of affected individuals do not have any additional major comorbidities affecting the musculoskeletal system [3], which allows them to achieve good daily performance results when using a prosthesis [4]. However, AKA contributes to gait asymmetry [5]. Increased levels of gait asymmetry may not merely adversely affect the joints; those individuals who undergo a unilateral amputation are also predisposed to flexion contracture of the thigh residual limb (especially in a short residual limb), osteoarthritis of the contralateral limb [6], and more frequent development of lower back pain (LBP) [7, 8]. Therefore, a significant number of potential health-related outcomes resulting from an asymmetrical gait are worthy of in-depth examination.

Increased levels of gait asymmetry usually occur in clinical populations such as individuals with lower limb amputations. These asymmetries are often considered to be a consequence of a pathology in which a patient compensates for movements to avoid pain or because of the desire to align the asymmetrical gait pattern [9]. According to research, the degree of gait symmetry of individuals who underwent unilateral AKA is influenced by the cause of amputation [10, 11], level of physical activity [12, 13] or type of prosthesis [14,15,16,17]. The key elements of the lower limb prosthesis are a well-fitted prosthetic socket and prosthetic knee joint module. With AKA, correctly choosing the prosthetic knee joint is important for safety and quality of movement [18]. The prosthetic limb may perform significantly more eccentric hip flexor work than the normal limb because the lack of prosthetic knee flexion is associated with a lack of hip flexion in an early stance, especially when compared to the intact limb [19].

The gait analysis of individuals with a unilateral AKA indicated changes in the range of movement mainly in pelvic tilt, pelvic obliquity, and hip abduction [20]. Knee prosthetics are currently described as mechanical modules (MechPK) or microprocessor-controlled modules (MicPK). The main difference in operation between MechPK and MicPK lies in the control of the swing and stance phases. MechPK only provides swing or stance phase control with manual locking, constant friction, weight-activated friction, and mechanisms that could increase stability in the stance phase. MicPK, on the other hand, is equipped with load and position sensors, which enables a more complete imitation of knee movements. In available studies, individuals with MicPK were characterized by a smaller asymmetry of thigh movements in the hip and knee joints [21, 22] and better balance [23]. However, there are reports suggesting that MechPK predominates over MicPK among patients [24].

Although some scalar indicators are a measure of space–time symmetry, they still do not provide a comprehensive assessment of overall gait symmetry. It has been observed that asymmetry of the step length (using the symmetry index, SI the relative difference between sides) for individuals who underwent an amputation differed between participants, and the results suggested symmetry for some even though their overall gait was asymmetrical [25, 26]. Therefore, a single measurement cannot provide extensive information on gait symmetry because asymmetries in the course of angle variations (angle-time characteristics), from which those metrics are derived, are not well described by numerical values. To accurately assess gait symmetry, the behavior of a whole limb (or a limb in a prosthesis) throughout the entire gait cycle should be considered. This suggests the need for more complete methods of measuring symmetry.

Conducting research that aims at detecting areas of increased differences between lower limbs in the gait of individuals who underwent AKA should optimize gait re-education. The limb amputation protocols for rehabilitation stages very often assume that supplies are selected based on progress in physiotherapy. However, this decision is usually based on the experience of the treatment team rather than on empirical data. We adopted the hypothesis that individuals equipped with MicPK knee will have better gait symmetry results than people with MechPK. The available literature offers only a limited number of studies presenting the results of such analysis [27].

Therefore, this study aimed at evaluating kinematic parameters of pelvic and hip joints in gait and looking for differentiating areas with an increased level of gait cycle asymmetry, compared to a normal gait, in individuals who underwent AKA and were equipped with MicPK or MechPK. The following research questions to be addressed: (1) Is the range of pelvic and thigh movements in the hip joint in the gait cycle of people equipped with MicPK different from that in individuals with MechPK? (2) Are there any characteristic adaptive variations between using prostheses with MicPK and MechPK expressed in the magnitude of gait cycle asymmetry? (3) Are there any differences in the gait cycle compared to the normal values (gait of healthy individuals) that support the use of MicPK in individuals who underwent unilateral AKA?

Methods

Study design, settings, and participants

Twenty-eight individuals participated in this cross-sectional study; 14 were assigned to an experimental group (individuals who underwent unilateral AKA), and the other 14 were healthy participants as a control group (Table 1). The control group showed no differences in terms of somatic structure and also representative of the population of middle-aged women and men.

Table 1 Demographic characteristics of study participants (mean ± SD)

This study was approved by the Bioethics Committee at the Medical University in Wroclaw (approval no. KB–232/2016). Informed verbal and written consent to take part in this study was obtained from each participant. The study was carried out in accordance with the tenets of the Declaration of Helsinki and Good Clinical Practice guidelines. This study adhered to the STROBE (Strengthening the Reporting of Observational Studies in Epidemiology) guidelines for observational studies.

Qualification procedure

Inclusion criteria were as following: (1) being at least 18 years old on the day of the examination, (2) having a reason other than vascular for the amputation, (3) using current prosthetic equipment on a daily basis for at least 6 months prior, and (4) the ability to walk at least 30 m without any help (Medicare functional classification level 3 or 4).

Exclusion criteria were as following: (1) a history of musculoskeletal disorders that cause pain, (2) fatigue, (3) a reduced range of lower limb movement or loss of coordination, (4) a dysfunction of the neuromuscular, cardiovascular or respiratory systems, and (5) cognitive impairments. Seven individuals equipped with single-axis hydraulic C-Leg were assigned to the MicPK group. The remainder of the participants were equipped with hydraulic polycentric knee unit 3R80 or 3R95 (without the microprocessor control) and assigned to the MechPK group. Other clinical and demographic characteristics are presented in Tables 1 and 2.

Table 2 Clinical characteristics in above-knee amputees

Measurement conditions

The study was conducted in a certified Biomechanical Analyses Laboratory of the Department of Biomechanics at the Academy of Physical Education in Wroclaw (Poland) between June and September 2019. The movement task was recorded using the optoelectronic SMART-E system (BTS Bioengineering, Milan, Italy) following the standard Davis protocol (the Newington model). The system consisted of six infrared (IR) cameras detecting reflecting markers with a sampling frequency of 120 Hz. Twenty-two markers were placed above the following anatomical points on participants’ skin: C7, the acromion processes, sacrum, anterior superior iliac spines (ASIS), greater trochanters, thighs (midpoint between the greater trochanter and lateral head of the femur), knees (lateral head of the femur and fibula), shank (midpoint between the lateral head of the fibula and the lateral malleolus), lateral malleoli, second metatarsal heads, and heels.

The measuring station was equipped with Smart Capture software for data recording, Smart Tracker for tracking the filming markers, and Smart Analyzer for data analysis and processing. The computer’s central processing unit contained a video controller (System VIX) and three network hubs (one 32-channel analog hub and two digital hubs, each with four communication ports). The IR cameras were rigidly fixed on the lab walls via frames [28].

Measurement procedure

Participants were asked to walk a distance of approximately 6 m both ways three times at their preferred walking speed, which allowed for an isolation of eight cycles for both lower limbs (measured together).

The individual biomechanical assessment included measuring the angle-time relationship characterizing the range of motion (ROM) in the lower limbs’ joints, especially (1) the pelvic obliquity angle (up/down)—rotation of the mediolateral axis out of the horizontal plane, where a positive value (up) corresponds to anterior superior iliac spine (ASIS) and posterior superior iliac spine (PSIS) markers that are higher than the corresponding markers on the contralateral side; (2) the pelvic tilt angle (up/down)—anterior/posterior rotation around the mediolateral axis, where a positive value (up) corresponds to the normal situation in which the PSIS is higher than the ASIS; (3) pelvic rotation angle (internal/external)—rotation of the mediolateral axis around the vertical axis; (4) the hip ad/abduction angle—rotation of the proximal–distal axis out of the sagittal anatomical plane; (5) the hip flexion–extension angle—rotation of the proximal–distal axis around the mediolateral axis, where a positive (flexion) angle value corresponds to the knee being in front of the body; (6) the hip rotation angle (internal/external)—rotation around the proximal–distal axis; and (7) the knee flexion–extension angle—rotation of the proximal–distal axis around the mediolateral axis, where a positive angle corresponds to a flexed knee [29, 30].

The following parameters were extracted: mean value, peak positive and negative values (Peak + and Peak −), and ROM calculated from the angle-time characteristics to statistically compare the differences. The gait cycle began when a limb first contacted the ground (at 0% GC). Based on the mathematical formula of SI [31], we propose a method of quantifying the gait symmetry with a help of the symmetry function (SF), which is an integral measure of absolute differences in time-standardized signals between the sides throughout the whole cycle of motion variation [32].

\({\text{SF}}(t) = 2 \cdot \frac{{x_{{{\text{un}}}} \left( t \right) - x_{{{\text{in}}}} \left( t \right)}}{{{\text{Range}}\left( {x_{{{\text{un}}}} (t)} \right) + {\text{Range}}\left( {x_{{{\text{in}}}} (t)} \right)}}\),

where xun and xin are any variables in function of time (time series) for the uninvolved and involved sides and Range is a range of variation for those characteristics.

Because a function constructed this way can combine variables of different physical dimensions, the difference signals have been normalized to the appropriate average range of their variability. SF functions have been grouped according to the relevant plane of movement (or movement degree of freedom), as in the case of angular waveforms, to track the process of physiotherapy in gait re-education.

Statistical analysis

Descriptive statistics such as Mean, Min, Max, and Range were carried out for (a) angle variation, (b) symmetry function (SF) between the left and right side, (c) SF between the TFA.UN limb and a intact limb of the control group (Norm), as well as, (d) SF between TFA.IN limb and a intact limb of the control group (Norm) based on the type of the knee prosthesis (C-Leg / Mech). The non-parametric methods were used for comparison the knee prosthesis (C-Leg / Mech) and parametric test were used for group comparison (i.e., AKA vs. healthy). Collected parameters were tested for statistically significant differences using a normality test (Kolmogorov–Smirnov test) as well as a test for the equality of group variances (Brown-Forsythe test). The significance of differences between groups was measured using one-way ANOVA with Bonferroni post-hoc test (for variables with distribution close to the normal distribution) and non-parametric Mann–Whitney U test with continuity correction (for variables that deviate from the normal distribution) with the assumed significance level alpha = 0.05. All statistical procedures were carried out using Statistica 13.1 software (TIBCO Software Inc., Palo Alto, CA, USA).

Results

The main speed (and ± 1 standard deviation) for the MicPK group was 0.97 ± 0.10 (m/s) and 0.91 ± 0.08 (m/s) for the MechPK group. There was no statistically significant difference between the speeds of test groups. The course of change in the angle of the pelvic movement in the gait cycle on each side was characterized by similar values in both the MechPK and MicPK groups (Table 3).

Table 3 The mean, minimum, maximum, and range of angles characterizing pelvic movements in frontal (POBLI), sagittal (PTILT), and transverse (PROT) plane as well as hip joint movements in frontal (HPAA), sagittal (HPFE), and transverse (HPROT) plane and knee joint movements sagittal plane (KFE) of an amputated limb (in) and an intact limb (un)

The characteristics of thigh movements in the hip joint of the intact limb for the MechPK group significantly differed for the frontal plane when compared with values recorded for the MicPK group (differences in mean 115%, p < 0.00; max 113%, p < 0.056; min 72%, p < 0.00). The thigh of the amputated limb of individuals from the MicPK group was adducted throughout the whole gait cycle. The ranges of motion for both groups and limbs in the sagittal plane were similar. In the transverse plane, the thigh of the amputated limb of those in the MicPK group was rotated externally at the end of the stance phase. For the MechPK group, on the side of the amputated limb, the highest values of the external rotation occurred at the end of the swing phase (differences in mean 70%, p < 0.554; max 65%, p < 0.313; min 8%, p < 1.232) (Fig. 1).

Fig. 1
figure 1

Graphs presenting the angle variation for the function duration of the gait cycle for a prosthetic limb (solid line) and an intact limb (dashed line) for the MicPK group (on the top) and MechPK group (on the bottom) characterizing pelvic movements in frontal (POBLI), sagittal (PTILT), and transverse (PROT) plane as well as hip joint movements in frontal (HPAA), sagittal (HPFE), and transverse (HPROT) plane

Applying the SF equation for pelvic movements on the side of the amputated limb in the frontal plane determined that the change of the asymmetry values did not significantly differ for either group (Table 4).

Table 4 The mean, minimum, maximum, and range values of the symmetry function characterizing pelvic movements in frontal (POBLI), sagittal (PTILT), and transverse (PROT) plane as well as hip joint movements in frontal (HPAA), sagittal (HPFE), and transverse (HPROT) plane and knee joint movements sagittal plane (KFE)

The highest asymmetry values of pelvic movements in the sagittal plane occurred in the stance phase for both groups. The pelvis was tilted forward for longer in the MechPK group. The asymmetry of pelvic movements in the transverse plane occurred throughout the whole gait cycle for both groups. The pelvis was turned toward the contralateral limb (differences in mean 83%, p < 0.376; max 126%, p < 0.387; min 73%, p < 0.354) (Fig. 2).

Fig. 2
figure 2

Graphs presenting variability of the axial symmetry function in relation to the duration of the gait cycle for the amputated limb (solid line) ± average standard deviation (dashed line) for MicPK group (on the top) and MechPK group (on the bottom) characterizing pelvic movements in frontal (POBLI), sagittal (PTILT), and transverse (PROT) plane as well as hip joint movements in frontal (HPAA), sagittal (HPFE), and transverse (HPROT) plane

Statistically significant differences in asymmetry of thigh movements in the hip joint of the amputated limb occurred in the frontal plane. In the MicPK group, the thigh was characterized by lower adduction asymmetry values and higher asymmetry values in the transverse plane and was rotated internally throughout the whole gait cycle. In the MechPK group, the thigh was more abducted throughout the whole gait cycle and less internally rotated. The ranges of asymmetry of thigh movements in the sagittal plane were similar, and the difference between groups in the size of the asymmetry occurred mainly in the swing phase (differences in mean 106%, p < 0.03; max 71%, p < 0.00; min 172%, p < 0.00) (Fig. 3).

Fig. 3
figure 3

Graphs presenting variability of the symmetry function (relative to the norm) in relation to the duration of the gait cycle for the amputated limb (solid line) ± average standard deviation (dashed line) for the MicPK group (on the top) and MechPK (on the bottom) characterizing pelvic movements in frontal (POBLI), sagittal (PTILT), and transverse (PROT) plane as well as hip joint movements in frontal (HPAA), sagittal (HPFE), and transverse (HPROT) plane

On the side of the amputated limb, the course of the range of pelvic movement asymmetry variability in the frontal plane of the gait cycle presented similar values for both groups (Table 5). The range of asymmetry compared to the movements of healthy individuals was the highest in the middle stance and swing phases when the pelvis on the side of the amputated limb was lowered in both groups (differences in mean 89%, p < 0.329). In the sagittal plane, for both MechPK and MicPK, the pelvis on the side of the amputated limb was tilted forward throughout the entire gait cycle. Higher asymmetry values were reported in the MechPK group, where the level of asymmetry ranged from 300 to 400%, whereas in the MicPK group, it ranged from 110 to 200% (differences in mean 47%, p < 0.353; max 38%, p < 0.402; min 64%, p < 0.352). In the transverse plane, the pelvis was rotated toward the contralateral direction at the beginning of the stance phase for both groups. The size of SF in the gait cycle for the amputated limb did not significantly differ between the groups.

Table 5 The mean, minimum, maximum, and range values of the variability of the symmetry function (relative to the norm) characterizing pelvic movements in frontal (POBLI), sagittal (PTILT), and transverse (PROT) plane as well as hip joint movements in frontal (HPAA), sagittal (HPFE), and transverse (HPROT) plane and knee joint movements sagittal plane (KEF) of an amputated limb (in) and an intact limb (un)

The level of thigh movement asymmetry in the hip joint of an amputated limb in the frontal plane differed between the measured groups in the stance phase. In the MicPK group, the thigh was positioned in greater abduction at the beginning of the stance phase than it was in the MechPK group. In the middle of the stance phase, the level of asymmetry was greater for individuals in the MechPK group. Moreover, in this group, the thigh in the hip joint was more adducted. The level of asymmetry compared to healthy individuals reached 100% (differences in mean 67%, p < 0.124; max 69%, p < 0.125; min 65%, p < 0.123). In the sagittal plane, the range of flexion and extension movement asymmetry of the amputated limb was similar and only slightly different from the physiological gait. In the transverse plane, the thigh of the amputated limb was rotated externally at the beginning of the stance phase. The level of asymmetry compared to healthy individuals reached 50% (Fig. 4).

Fig. 4
figure 4

Graphs presenting variability of the symmetry function (relative to the norm) in relation to the duration of the gait cycle for the intact limb (solid line) ± average standard deviation (dashed line) for the MicPK group (on the top) and MechPK (on the bottom) characterizing pelvic movements in frontal (POBLI), sagittal (PTILT), and transverse (PROT) plane as well as hip joint movements in frontal (HPAA), sagittal (HPFE), and transverse (HPROT) plane

The variability of angular values after applying the SF equations for pelvic movements of the intact limb in the frontal plane in the stance phase was similar for both groups. The largest differences compared to pelvic movements in the frontal plane for healthy individuals occurred in the swing phase, when the pelvis was positioned nearly 50% higher. In the sagittal plane, similar to pelvic movement asymmetry on the side of the amputated limb, the pelvis was tilted forward on the side of the intact limb throughout the whole gait cycle (differences in mean 36%, p < 0.383; max 27%, p < 0.404; min 53%, p < 0.372). The level of asymmetry was similar to that of the amputated limb, higher for the MechPK group (200–300%) than for the MicPK group (100–200%). Pelvic movements in the transverse plane were characterized by the highest asymmetry values in the final stage of the stance phase. The pelvis was rotated toward the contralateral limb. The size of the SF in the gait cycle for the amputated limb did not significantly differ between groups.

Statistically significant between-group differences in the asymmetry of thigh movement range in the hip joint of the intact limb were found in the frontal plane. The intact limb in the middle of the stance phase was more adducted in the MicPK group and less adducted in the MechPK group (differences in mean 180%, p < 0.000; max 63%, p < 0,000; min 65%, p < 0.000). In the sagittal plane, the range of asymmetry of the flexion and extension of the thigh of the intact limb was similar and only slightly different from the physiological gait. In the transverse plane, higher asymmetry values were observed for individuals from the MechPK group. The thigh was more externally rotated at the end of the swing stage and more internally rotated in the swing phase. The level of asymmetry compared to the gait of healthy individuals reached 50% (differences in mean 115%, p < 0.232; max 62%, p < 0.26; min 50, p < 0.154).

Discussion

Starting the rehabilitation of individuals who underwent lower limb amputations is difficult. Amputation for reasons other than vascular is often done to save the patient’s life. Individuals then entrust considerable hope in effective and optimal prosthetics, in which they see a possibility of returning to their previous lifestyle. Choosing the right type of prosthesis is an important and key stage, therefore this research might help to making the appropriate choice of prosthetic.

Humans’ symmetrical gait pattern is characterized by an almost identical course on both sides of the body. This symmetry usually deteriorates into a pathological gait due to pain or disturbances in the normal function of the locomotive system resulting from AKA. Over time, the persistence of the asymmetrical gait pattern can lead to other musculoskeletal dysfunctions. Based on this assumption, we conducted research aimed at confirming the existence of differences in gait asymmetry in the population after AKA, in relation to the type of prosthetic supplies. We also focused on the assessment of effectiveness and the feasibility of using the SF tool to diagnose gait symmetry.

Asymmetry might be investigated from the point of view of time-spatial parameters (e.g., uneven walking lengths) and angular parameters (e.g., a difference in the ROM between the amputated and sound limbs). The differences in ROM varied between experimental subgroups in the frontal plane in the hip joint of the intact limb. Seroussi et al. [19] presented similar results for the ankle, knee, and hip: 5.6, 6.3, and 5.5 degrees, respectively, for the prosthetic limb and 6.6, 4.7, and 7.3 degrees, respectively, for the intact limb. The range of hip movements in the sagittal plane in individuals from the experimental group was similar to the normal gait value. Statistical analysis did not confirm other differences in the gait cycle between the MechPK and MicPK groups. Kaufman et al. [16] also reported that the range of movements in the sagittal plane in both amputated and intact limbs was similar in compared groups of individuals after AKA equipped with MechPK and MicPK. These results allow us to conclude that both prosthetic modules correctly imitate physiological knee movements, but the time of the intensity of asymmetry in the gait cycle differs.

The proposed method identifies where in the gait cycle the asymmetry is less or more pronounced. The suggested measure of symmetry measures not only the overall symmetry in each component of movement but also explains whether the asymmetry is caused by a shift in time or difference in values. The thigh of individuals in the MechPK group was abducted throughout the whole gait cycle and externally rotated. The thigh of those in the MicPK group was also abducted, but the degree of asymmetry was lower. A higher degree of asymmetry occurred in external rotation movements. These results may indicate a different degree of adaptation for people who underwent AKA, resulting from a difference in knee prosthetic modules’ performance. Kaufman et al. [21] state that the gait of individuals equipped with MicPK is similar to the correct one. However, our research shows that the gait of people with MicPK might be more similar to the correct gait in the frontal plane, whereas the degree of asymmetry in the transverse plane is greater than that of people with MechPK. This also influences an uneven load distribution, which results from an asymmetrical gait pattern regardless of module type [33]. Luetmer et al. [8] and Devan et al. [34] state that the type of prosthetic does not reduce the frequency of LBP, confirming this idea.

Unlike the known scalar-based measures of symmetry, a function-based measure quantifies symmetry over the entire gait cycle and, in certain cases, at numerous levels of analysis. Time waveforms can be based on kinematic or kinetic variables, ground reaction forces, or muscle activity patterns that are available as a result of motion analysis [32]. The highest values of pelvic movement asymmetry compared to healthy individuals occurred in the sagittal plane [35]. Research shows that this result may be influenced by the type of prosthetic socket [36]. Participants in the experimental group used an anatomical prosthetic socket, in which the thigh residual limb is supported on the medial wall or on all walls; one example is the Marlo Anatomical Socket (MAS) type. According to Rabuffetti et al. [37], relative to healthy individuals, those who underwent AKA were characterized by a reduction of extension in the hip joint on the side of the amputated limb of 7.5 ± 4.4 degrees, an increase in pelvic inclination of 8.2 ± 3.6 degrees, and increase in hip joint flexion on the side of the intact limb of 13.8 ± 7.4 degrees. Participants in the experimental group had a different pattern of asymmetry regardless of the prosthetic socket. The efficiency of the use of a lower limb prosthesis largely depends on the length the thigh stump, muscle strength and flexibility of the surrounding tissues. Previous studies have confirmed the impact of the generated moment of force on the gait pattern of individuals after AKA [32, 38]. Therefore, the differences in the use of lower limb prostheses can be observed in the efficiency of the thigh residual limb.

The number of publications on gait asymmetry in clinical trials is increasing due to the potential implications. Individuals after AKA in a gait re-education program can use a simple measure of symmetry that can be easily and quickly interpreted. This suggestion is confirmed by research showing that learning and maintaining complex motor skills (such as a more symmetrical gait pattern) is reinforced by external feedback associated with the overall assessment of locomotion performance rather than specific movements. In this study, different strategies were observed for a lower limb prosthesis depending on the type of knee module.

Study limitations

The present study has some potential limitations. First, the research was carried out in the clinical conditions of only one center. Second, the size of the AKA group was not large and was insufficiently representative to generalize changes in differences in the range of movements in individuals equipped with MechPK and MicPK to the whole population of individuals after unilateral AKA. Finally, the research does not concentrate on the relationship between gait analysis outcomes and the physical characteristics of the participants.

Practical implications

When choosing the final prostheses for individuals with unilateral AKA, it is important to first assess the condition and functional efficiency of the thigh residual limb. Furthermore, less technically advanced prostheses are similar to the more advanced solutions in allowing for restoring a pattern of the gait cycle. The type of prosthetic equipment should be considered, due to differences in gait movement compensation, especially in the frontal plane.

Conclusions

The results of this study allowed for the following conclusions to be drawn. MicPK reduces the asymmetry values when compared to the movements of healthy individuals, especially the movements in the sagittal plane and pelvic rotations. Participants equipped with MechPK and MicPK develop a different type of gait in the prosthesis, expressed by changing the size of the asymmetry of the pelvic and thigh movements in the hip joint. Furthermore, in the experimental group, the kinematic variables of pelvic and thigh movements in the hip joint differ only in movements in the frontal plane. Individuals who underwent unilateral AKA can gain less from using an MicPK lower limb prosthesis than was anticipated. Our observations show, both in people with MicPK and MechPK protheses, the degree of adduction of the thigh in the hip joint during the support phase is increased. This results in the need for greater involvement of the abduction muscles of the hip joint (mainly the gluteus medius muscle). This study shows that the choice of module in the final prosthesis should be based on the size of changes in the thigh stump and depend on the ability to control the prosthesis in frontal plane individuals with short limb stumps.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

AKA:

Above-knee amputation

ASIS:

Anterior superior iliac spines

IR:

Infrared

MAS:

Marlo anatomical socket

MechPK:

Mechanical modules

MicPK:

Microprocessor modules

PSIS:

Posterior superior iliac spine

ROM:

Range of motion

SF:

Symmetry function

STROBE:

Strengthening the reporting of observational studies in epidemiology guidelines

References

  1. George J, Navale SM, Nageeb EM, Curtis GL, Klika AK, Barsoum WK, Mont MA, Higuera CA. Etiology of above-knee amputations in the united states: is periprosthetic joint infection an emerging cause? Clin Orthop Relat Res. 2018;476(10):1951–60.

    Article  Google Scholar 

  2. Low EE, Inkellis E, Morshed S. Complications and revision amputation following trauma-related lower limb loss. Injury. 2017;48(2):364–70.

    Article  Google Scholar 

  3. Gailey RS, Roach KE, Applegate EB, Cho B, Cunniffe B, Licht S, Maguire M, Nash MS. The amputee mobility predictor: an instrument to assess determinants of the lower-limb amputee’s ability to ambulate. Arch Phys Med Rehabil. 2002;83(5):613–27.

    Article  Google Scholar 

  4. Mundell BF, Kremers HM, Visscher S, Hoppe KM, Kaufman KR. Predictors of receiving a prosthesis for adults with above-knee amputations in a well-defined population. PM R. 2016;8(8):730–7.

    Article  Google Scholar 

  5. Ray SF, Wurdeman SR, Takahashi KZ. Prosthetic energy return during walking increases after 3 weeks of adaptation to a new device. J Neuroeng Rehabil. 2018;15(1).

  6. Briem K, Snyder-Mackler L. Proximal gait adaptations in medial knee OA. J Orthop Res. 2009;27(1):78–83.

    Article  Google Scholar 

  7. Jayaraman C, Hoppe-Ludwig S, Deems-Dluhy S, McGuire M, Mummidisetty C, Siegal R, Naef A, Lawson BE, Goldfarb M, Gordon KE, Jayaraman A. Impact of powered knee-ankle prosthesis on low back muscle mechanics in transfemoral amputees: a case series. Neurosci Front. 2018. https://0-doi-org.brum.beds.ac.uk/10.3389/fnins.2018.00134.

    Article  Google Scholar 

  8. Luetmer M, Mundell B, Kremers HM, Visscher S, Hoppe KM, Kaufman KR. Low back pain in adults with transfemoral amputation: a retrospective population-based study. PMR. 2019;11(9):926–33.

    Article  Google Scholar 

  9. Narang YS, Arelekatti VNM, Winter AG. The effects of prosthesis inertial properties on prosthetic knee moment and hip energetics required to achieve able-bodied kinematics. IEEE Trans Neural Syst Rehabil Eng. 2016;24(7):754–63.

    Article  Google Scholar 

  10. Marina N, Hudson MM, Jones KE, Mulrooney DA, Avedian R, Donaldson SS, Popat R, West DW, Fisher P, Leisenring W, Stovall M, Robison LL, Ness KK. Changes in health status among aging survivors of pediatric upper and lower extremity sarcoma: a report from the childhood cancer survivor study. Arch Phys Med Rehabil. 2013;94(6):1062–73.

    Article  Google Scholar 

  11. Jarvis HL, Bennett AN, Twiste M, Phillip RD, Etherington J, Baker R. Temporal spatial and metabolic measures of walking in highly functional individuals with lower limb amputations. Arch Phys Med Rehabil. 2017;98(7):1389–99.

    Article  Google Scholar 

  12. Kowal M, Rutkowska-Kucharska A. Muscle torque of the hip joint flexors and extensors in physically active and inactive amputees. Hum Kinet Biomed. 2014. https://0-doi-org.brum.beds.ac.uk/10.2478/bhk-2014-0011.

    Article  Google Scholar 

  13. de Araújo HN, Mendes FAS, Fortes CE, Borin G, Garcia PA, Macedo OG, Marães VRFS, Durigan JLQ. Dynamic and static postural control in volleyball players with transfemoral amputation. Rev Bras Med do Esporte. 2019;25(1):58–62.

    Article  Google Scholar 

  14. Chen C, Hanson M, Chaturvedi R, Mattke S, Hillestad R, Liu HH. Economic benefits of microprocessor controlled prosthetic knees: a modeling study. J Neuroeng Rehabil. 2018;15(S1):62.

    Article  Google Scholar 

  15. Lura DJ, Wernke MM, Carey SL, Kahle JT, Miro RM, Highsmith MJ. Differences in knee flexion between the Genium and C-Leg microprocessor knees while walking on level ground and ramps. Clin Biomech. 2015;30(2):175–81.

    Article  Google Scholar 

  16. Kaufman KR, Frittoli S, Frigo CA. Gait asymmetry of transfemoral amputees using mechanical and microprocessor-controlled prosthetic knees. Clin Biomech. 2012;27(5):460–5.

    Article  Google Scholar 

  17. Taheri AR, Karimi MT. Evaluation of the gait performance of above-knee amputees while walking with 3R20 and 3R15 knee joints. J Res Med Sci. 2012;17(3):6–11.

    Google Scholar 

  18. Hafner BJ, Willingham LL, Buell NC, Allyn KJ, Smith DG. Evaluation of function, performance, and preference as transfemoral amputees transition from mechanical to microprocessor control of the prosthetic knee. Arch Phys Med Rehabil. 2007;88(2):207–17.

    Article  Google Scholar 

  19. Seroussi RE, Gitter A, Czerniecki JM, Weaver K. Mechanical work adaptations of above-knee amputee ambulation. Arch Phys Med Rehabil. 1996;77(11):1209–14.

    Article  CAS  Google Scholar 

  20. Bae TS, Choi K, Hong D, Mun M. Dynamic analysis of above-knee amputee gait. Clin Biomech. 2007;22(5):557–66.

    Article  Google Scholar 

  21. Kaufman KR, Levine JA, Brey RH, Iverson BK, McCrady SK, Padgett DJ, Joyner MJ. Gait and balance of transfemoral amputees using passive mechanical and microprocessor-controlled prosthetic knees. Gait Posture. 2007;26(4):489–93.

    Article  CAS  Google Scholar 

  22. Bellmann M, Schmalz T, Ludwigs E, Blumentritt S. Immediate effects of a new microprocessor-controlled prosthetic knee joint: a comparative biomechanical evaluation. Arch Phys Med Rehabil. 2012;93(3):541–9.

    Article  Google Scholar 

  23. Kaufman KR, Levine JA, Brey RH, McCrady SK, Padgett DJ, Joyner MJ. Energy expenditure and activity of transfemoral amputees using mechanical and microprocessor-controlled prosthetic knees. Arch Phys Med Rehabil. 2008;89(7):1380–5.

    Article  Google Scholar 

  24. Segal AD, Orendurff MS, Klute GK, McDowell ML, Pecoraro JA, Shofer J, Czerniecki JM. Kinematic and kinetic comparisons of transfemoral amputee gait using C-Leg® and Mauch SNS® prosthetic knees. J Rehabil Res Dev. 2006;43(7):857–70.

    Article  Google Scholar 

  25. Błazkiewicz M, Wiszomirska I, Wit A. Comparison of four methods of calculating the symmetry of spatial-temporal parameters of gait. Acta Bioeng Biomech. 2014;16(1):29–35.

    PubMed  Google Scholar 

  26. Jochymczyk-Woźniak K, Nowakowska-Lipiec K, Zadoń H, Wolny S, Gzik M, Gorwa J, Michnik R. Gait kinematics index, global symmetry index and gait deviations profile: concept of a new comprehensive method of gait pathology evaluation. Acta Bioeng Biomech. 2020;22(4):61–73.

    Google Scholar 

  27. Lencioni T, Carpinella I, Rabuffetti M, Marzegan A, Ferrarin M. Human kinematic, kinetic and EMG data during different walking and stair ascending and descending tasks. Sci Data. 2019;6:309.

    Article  Google Scholar 

  28. Roerdink M, Roeles S, van der Pas SCH, Bosboom O, Beek PJ. Evaluating asymmetry in prosthetic gait with step-length asymmetry alone is flawed. Gait Posture. 2012;35(3):446–51.

    Article  Google Scholar 

  29. Wu G, Cavanagh PR. ISB recommendations for standardization in the reporting of kinematic data. J Biomech. 1995;28(10):1257–61.

    Article  CAS  Google Scholar 

  30. Wu G, Siegler S, Allard P, Kirtley C, Leardini A, Rosenbaum D, Whittle M, D’Lima DD, Cristofolini L, Witte H, Schmid O, Stokes I. ISB Recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion—Part I: ankle, hip, and spine. J Biomech. 2002;35(4):543–8.

    Article  Google Scholar 

  31. Robinson RO, Herzog W, Nigg BM. Use of force platform variables to quantify the effects of chiropractic manipulation on gait symmetry. J Manipulative Physiol Ther. 1987;10(4):172–6.

    CAS  PubMed  Google Scholar 

  32. Rutkowska-Kucharska A, Kowal M, Winiarski S. Relationship between asymmetry of gait and muscle torque in patients after unilateral transfemoral amputation. Appl Bionics Biomech. 2018. https://0-doi-org.brum.beds.ac.uk/10.1155/2018/5190816.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Welke B, Jakubowitz E, Seehaus F, Daniilidis K, Timpner M, Tremer N, Hurschler C, Schwarze M. The prevalence of osteoarthritis: higher risk after transfemoral amputation?—a database analysis with 1569 amputees and matched controls. PLoS ONE. 2019. https://0-doi-org.brum.beds.ac.uk/10.1371/journal.pone.0210868.

    Article  PubMed  PubMed Central  Google Scholar 

  34. Devan H, Hendrick P, Ribeiro DC, Hale L, Carman A. Asymmetrical movements of the lumbopelvic region: is this a potential mechanism for low back pain in people with lower limb amputation? Med Hypotheses. 2014;82(1):77–85.

    Article  Google Scholar 

  35. Tranberg R, Zügner R, Kärrholm J. Improvements in hip- and pelvic motion for patients with osseointegrated trans-femoral prostheses. Gait Posture. 2011;33(2):165–8.

    Article  CAS  Google Scholar 

  36. Fatone S, Dillon M, Stine R, Tillges R. Coronal plane socket stability during gait in persons with transfemoral amputation: pilot study. J Rehabil Res Dev. 2014;51(8):1217–28.

    Article  Google Scholar 

  37. Rabuffetti M, Recalcati M, Ferrarin M. Trans-femoral amputee gait: socket-pelvis constraints and compensation strategies. Prosthet Orthot Int. 2005;29(2):183–92.

    Article  CAS  Google Scholar 

  38. Croisier JL, Maertens de Noordhout B, Maqueta D, Camus G, Hac S, Feron F, De Lamotte O, Crielaard JM. Isokinetic evaluation of hip strength muscle groups in unilateral lower limb amputees. Isokinet Exerc Sci. 2001;9(4):163–9.

    Article  Google Scholar 

Download references

Acknowledgements

The authors thank all the participants in present study for their participation and valuable feedbacks.

Funding

The present study was funded by the Ministry of Health in Poland as a statutory grant of the Wroclaw Medical University for maintaining research potential (no. SUB.E060.19.001).

Author information

Authors and Affiliations

Authors

Contributions

MK, SW, ARK, and AK designed the experiment, analysed and interpreted the data, and drafted the manuscript. WB contributed to study design, participant recruitment, data collection and data processing. MPB contributed to study design, participant recruitment, data collection and data processing as well as contributed to revision of manuscript. KW and EG helped conducting the trials and improving test protocol. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Ewa Gieysztor.

Ethics declarations

Ethics approval and consent to participate

This study was approved by the Bioethics Committee of the Wroclaw Medical University in Poland (no. KB-232/6016). Written informed consent was obtained from all participants.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kowal, M., Winiarski, S., Gieysztor, E. et al. Symmetry function in gait pattern analysis in patients after unilateral transfemoral amputation using a mechanical or microprocessor prosthetic knee. J NeuroEngineering Rehabil 18, 9 (2021). https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-021-00810-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-021-00810-w

Keywords