Skip to main content

Table 5 Comparison of the proposed instrument with other state-of-the-art wearable and wireless biopotential acquisition systems existing in the literature

From: A high-performance 8 nV/√Hz 8-channel wearable and wireless system for real-time monitoring of bioelectrical signals

Parameters

[43]

[44]

[45]

This work

Number of channels

32

8

4

5 (+ 3)

Type of biosignals

EMG

ECG, EMG

EEG, LFP

ExG, ECoG, LFP, ERNA, body position and movement, PV ectopic activity (+ amperometric/potentiometric)

Voltage gain

46 dB

≈ 28 dB (max)

≈ 54 dB

60 dB

Input voltage range

±5 mV

±100 mV

±1.15 mV

±2.3 mV

Maximum available bandwidth

10–500 Hz

0–250 Hz

1.5–100 Hz

0.5–500 Hz

Input impedance

1.3 GΩ (10 Hz)

≈ 620 MΩ (10 Hz)

47 kΩ (10 Hz)

≈ 8 GΩ (10 Hz)

Input referred noise

<  3 μV rms

0.2 μV rms (0–250 Hz)

≈ 0.606 μV rms (1.5–100 Hz)

0.169 μV rms (0.5–500 Hz)

CMRR

82 dB

110 dB

134 dB

Maximum resolution of the ADC

16 bits

24 bits

16 bits

24 bits

Maximum sampling frequency

2048 SPS

500 SPS (16 kSPS for wired)

500 SPS

1 kSPS (32 kSPS for wired)

Hours of continuous operation

5 (600 mAh 1-Cell LiPo battery)

13.6 (1700 mAh battery)

6 to 8 (CR1/3 N lithium ion button-cell battery)

8 (1 Ah lithium battery)

Wireless capability

Yes (Wi-Fi)

Yes (Bluetooth)

Yes (2.4 GHz MSK)

Yes (Zigbee)

Latency

12 ms

1 ms

Max. transmission range

22 m

10 m

3–5 m

5 m

Area (length × width)

10.2 cm2

21.7 cm2 (PCB area)

4.76 cm2

44 cm2 (PCB area)