Effect of high-frequency alternating current transcutaneous stimulation over muscle strength: a controlled pilot study

Abstract

Background

High-frequency alternating currents of greater than 1 kHz applied on peripheral nerves has been used in animal studies to produce a motor nerve block. It has been evidenced that frequencies higher than 5 kHz are necessary to produce a complete peripheral nerve block in primates, whose nerve thickness is more similar to humans. The aim of the study was to determine the effect on muscle strength after the application of a high-frequency stimulation at 5 and 10 kHz compared to sham stimulation in healthy volunteers.

Findings

Transcutaneous stimulation at 5 kHz, 10 kHz and sham stimulation were applied to eleven healthy volunteers over the ulnar and median nerves for 20 min. Maximal handgrip strength was measured before, during, immediately after the intervention, and 10 min after the end of intervention. The 10 kHz stimulation showed a lower handgrip strength during the intervention (28.1 N, SEM 3.9) when compared to 5 kHz (31.1 N, SEM 3.6; p < 0.001) and to sham stimulation (33.7 N, SEM 3.9; p < 0.001). Furthermore, only stimulation at 10 kHz decreased handgrip strength when compared to baseline.

Conclusions

These findings suggest high-frequency stimulation has an inhibitory effect over muscle strength. Future studies are required in patients that are characterized by motor hyperactive such as spasticity or tremors.

Clinical trial registration

NCT, NCT03169049. Registered on 30 May 2017

Background

Previous studies in animals have shown that high-frequency alternating current (HFAC) of greater than 1 kHz applied on exposed peripheral nerves can produce a motor nerve block (Bhadra and Kilgore [1, 2]). An in vivo study, [3] showed that frequencies higher than 5 kHz were able to block nerve conduction of motor fibers. One study [4] in non-injured subjects showed an incomplete block when transcutaneous HFAC applied to the radial nerve at 5 kHz increased somatosensory thresholds. It has been evidenced [5] that frequencies higher than 5 kHz are necessary to produce a complete peripheral nerve block in primates, whose nerve diameter is similar to humans, however, there is not any human study that apply HFAC transcutaneously with frequencies higher than 5 kHz. It is believed that the nerve conduction block produced by application of HFAC could be a useful tool for the treatment of patients with pain or with an exaggerated increase of nerve activity, such as hypertonia or spasms.

The purpose of this study was to determine the effects on maximal handgrip strength (MHS) of a non-invasive HFAC at 5 kHz and 10 kHz applied to the ulnar and median nerves in healthy subjects, compared to a sham stimulation.

Methods

A randomized, crossover, single-blinded, placebo-controlled trial was conducted in 11 healthy volunteers after signing the informed consent approved by the Local Ethics Committee. Participants received three randomized (www.randomizer.org) interventions (10 kHz, 5 kHz, and sham stimulation) with a washout period of 24-h. Moreover, four measurements were registered: before (0 min); during the intervention at 15 min from onset (15 min); immediately after intervention (20 min) and 10 min after the intervention had finished (30 min).

All interventions were applied for 20 min with two surface self-adhesive electrodes (ValuTrode, Axelgaard Manufacturing, USA) 5 cm × 5 cm, which were placed on the anterior face of the dominant forearm, over the ulnar and median nerves. The proximal electrode was fixed in the path of the ulnar nerve over the epitrochlea and the distal electrode was placed on the median nerve over the carpal tunnel. A stimulator (Myomed 932. Enraf Nonius, Netherlands) delivered a sinusoidal current without modulation, at a frequency of 10 kHz. Intensity was determined by a “strong but comfortable tingle” sensation, just below motor threshold. The intensity was gradually increased until a minimally visible contraction was observed and subsequently decreased until it disappeared, and this sensation remained throughout the session. To avoid habituation to the stimulus, participant were asked every 2 minutes to corroborate the perceived sensation, and the intensity was increased if requested [6, 7]. The same procedure was performed with the 5 kHz stimulation. The sham stimulation session was applied by progressively increasing the intensity of a non-connected channel. Participants were also blinded to the hypothesis of the study by being informed that in some cases the perceived sensation might be different because the intensity could be adjusted to below their sensitive threshold, with the possibility that the participant may or may not feel the current [4, 8].

Dynamometry is a reliable and objective method to quantify manual grip strength. The absence or decrease of muscle contraction is considered as an indicator of motor nerve block [8], so dynamometry can be used as an indirect measure to assess the block of motor fibers in humans. Handgrip strength is mediated by ulnar and median nerves and was measured with a handgrip dynamometer (Grip Strength Dynamometer, Tokyo). Three repetitions were registered, and the mean was taken as a value of the average maximal strength.

Statistical analysis was performed with the software “SPSS Statistics 22.0”. Due to the confirmed normal distribution of the data, a parametric test was adopted. Two-way repeated measures ANOVA (time and intervention factors) was performed with a Bonferroni post-hoc test. p < 0.05 was considered significant.

Results

Eleven subjects (4 males, 36%) with a mean age of 22 years (SEM 1.9) completed the study. The 10 kHz intervention showed a greater initial and final intensity (Initial: 23 mA, SEM 3.3; Final: 39.9 mA, SEM 4.0) than the 5 kHz intervention (Initial: 13.0 mA, SEM 1.4; Final: 21.5 mA, SEM 2.1), p < 0.01 and p < 0.001, respectively. Table 1 shows data of MHS measurement of each participant. No differences were reported for maximal handgrip strength before the interventions. Significant differences in the “time” factor (F(3,8) = 4.49; p = 0.04), the “intervention” factor (F(2,9) = 8.38; p = 0.009) and the “time-intervention” interaction (F(6,5) = 6,08; p = 0.03) were detected. Specifically, HFAC applied at 10 kHz evidenced a statistically significant decrease in MHS at 15 min during the intervention (28.1 N, SEM 3.9; p = 0.001) when compared to baseline (32.6 N, SEM 4.0) and to 10 min post-stimulation (31.4 N, SEM 3.7; p = 0.02). Sham stimulation and 5 kHz interventions did not show any statistical change from baseline. Between-group comparisons showed a lower handgrip strength with 10 kHz (28.1 N, SEM 3.9) when compared to 5 kHz (31.1 N, SEM 3.6; p = 0.001) and to sham intervention (33.7 N, SEM 3.9; p = 0.003) during the stimulation. Immediately after the intervention, a lower MSH was also found in the 10 kHz group (29.9 N, SEM 3.7) when comparing to 5 kHz (32.1 N, SEM 3.6; p = 0.02) and sham stimulation (32.2 N, SEM 3.8; p = 0.03) (Fig. 1).

Table 1 Maximal handgrip strength measurement of each participant throughout the experimental sessions
Fig. 1
figure 1

Stimulation effect on maximal handgrip strength. Sham stimulation (circle), 5 kHz (square), and 10 kHz (triangle). Data are represented as mean and standard error. * Indicates significantly different compared to sham stimulation (***p < 0.001; **p < 0.01). + Indicates a significant difference compared to 5 kHz stimulation (++p < 0.01). # Indicates significantly different from baseline (###p < 0.001)

Discussion

This is the first study that applies transcutaneously HFAC at 10 kHz over a human peripheral nerve and a motor nerve conduction block is suggested evidenced by a decrease of MSH. Although there is a high variability among participants, the mean difference of handgrip strength among temporal points have a low variability. This high variability is due to the fact that males have higher levels of handgrip strength than females [9]. The higher intensity needed for the 10 kHz compared to the 5 kHz intervention to achieve the same current perception was expected because higher frequencies have less resistance and less perception of the electrical current [10].

These results are in consonance with experimental animal studies which have shown a peripheral motor conduction block with frequencies between 3 and 10 kHz [11,12,13]. However, only one study in humans [4] has previously applied transcutaneous HFAC at 5 kHz, with similar effects than conventional TENS on somatosensory thresholds assessment. The higher effect of the 10 kHz current compared to the 5 kHz confirms a frequency-dependent effect suggested by a previous study where the relationship between nerve diameter and block frequency was determined [5]. A key point to note is that the decrease of MHS evidenced during stimulation is quickly reversible when the stimulation ends, as well as the nerve conduction block observed in animals [14]. This could have relevant clinical applications to reduce specific motor activity such as blocking specific nerves during disabling spams or tremors. However, these effects should be studied in subjects with disorders characterized by motor hyperactivity.

Study limitations

This is a pilot study that assesses handgrip strength as an indirect indicator of motor fiber conduction. However, it is necessary to confirm these results using direct measures such as neurophysiological tests applied in both, healthy and pathological subjects. The major challenge of using neurophysiological tests during the application of HFAC is the artefact evoked by the electrical stimulation.

Conclusions

High-frequency alternating current stimulation without modulation, at a frequency of 10 kHz applied over ulnar and median nerves of non-injured volunteers produces, during and immediately after the stimulation, a decrease of MHS when compared to HFAC at 5 kHz and sham stimulation. This reduction in MHS could suggest a block of motor nerve activity that could be useful in subjects with neurological disorders characterized by motor hyperactivity.

Abbreviations

ANOVA:

Analysis of variance

HFAC:

High-frequency alternating current

MSH:

Maximal handgrip strength

N:

Newtons

SEM:

Standard error mean

References

  1. Bhadra N, Kilgore KL. High-frequency electrical conduction block of mammalian peripheral motor nerve. Muscle Nerve. 2005;32(6):782–90.

    Article  Google Scholar 

  2. Tai C, Roppolo JR, de Groat WC. Response of external urethral sphincter to high frequency biphasic electrical stimulation of pudendal nerve. J Urol. 2005;174(2):782–6.

    Article  Google Scholar 

  3. Williamson RP, Andrews BJ. Localized electrical nerve blocking. IEEE Trans Biomed Eng. 2005;52(3):362–70.

    Article  Google Scholar 

  4. Avendano-Coy J, et al. Effect of unmodulated 5-kHz alternating currents versus transcutaneous electrical nerve stimulation on mechanical and thermal pain, tactile threshold, and peripheral nerve conduction: a double-blind, placebo-controlled crossover trial. Arch Phys Med Rehabil. 2017;98(5):888–95.

    Article  Google Scholar 

  5. Ackermann DM, et al. Electrical conduction block in large nerves: high-frequency current delivery in the nonhuman primate. Muscle Nerve. 2011;43(6):897–9.

    Article  Google Scholar 

  6. Serrano-Munoz D, et al. Intensity matters: therapist-dependent dose of spinal transcutaneous electrical nerve stimulation. PLoS One. 2017;12(12):e0189734.

    Article  Google Scholar 

  7. Aarskog R, et al. Is mechanical pain threshold after transcutaneous electrical nerve stimulation (TENS) increased locally and unilaterally? A randomized placebo-controlled trial in healthy subjects. Physiother Res Int. 2007;12(4):251–63.

    Article  Google Scholar 

  8. Claydon LS, et al. Alternating-frequency TENS effects on experimental pain in healthy human participants: a randomized placebo-controlled trial. Clin J Pain. 2013;29(6):533–9.

    Article  Google Scholar 

  9. Sisto SA, Dyson-Hudson T. Dynamometry testing in spinal cord injury. J Rehabil Res Dev. 2007;44(1):123–36.

    Article  Google Scholar 

  10. Beatti A, et al. Penetration and spread of interferential current in cutaneous, subcutaneous and muscle tissues. Physiotherapy. 2011;97(4):319–26.

    Article  Google Scholar 

  11. Yang G, et al. Post-stimulation block of frog sciatic nerve by high-frequency (kHz) biphasic stimulation. Med Biol Eng Comput. 2017;55(4):585–93.

    Article  Google Scholar 

  12. Kilgore KL, Bhadra N. Nerve conduction block utilising high-frequency alternating current. Med Biol Eng Comput. 2004;42(3):394–406.

    Article  CAS  Google Scholar 

  13. Ackermann DM, et al. Effect of bipolar cuff electrode design on block thresholds in high-frequency electrical neural conduction block. IEEE Trans Neural Syst Rehabil Eng. 2009;17(5):469–77.

    Article  Google Scholar 

  14. Ackermann DM, et al. Conduction block of peripheral nerve using high-frequency alternating currents delivered through an intrafascicular electrode. Muscle Nerve. 2010;41(1):117–9.

    Article  Google Scholar 

Download references

Funding

This work was supported by the “Instituto de Salud Carlos III” (ref PI17/00581).

Availability of data and materials

The data collected in this study are available from the corresponding author on reasonable request.

Author information

Authors and Affiliations

Authors

Contributions

DSM and CSM contributed in participant recruitment, data collection and drafting manuscript. JAC, JGS and JT conceiving study concept, results interpretation and supervised the whole project. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Juan Avendaño-Coy.

Ethics declarations

Ethics approval and consent to participate

The local ethical committee of the “Complejo Hospitalario de Toledo” gave approval to the study and warranted its accordance with the Declaration of Helsinki (number 158; 2-11-2017). Written informed consent was obtained from all participants prior to participation.

Consent for publication

Not applicable.

Competing interests

Avendaño-Coy, Taylor and Gómez-Soriano have a patent of a device which evokes electrical current between 10 kHz and 50 kHz, although this device has not used in this work. The remaining authors have no conflicts of interest.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Serrano-Muñoz, D., Avendaño-Coy, J., Simón-Martínez, C. et al. Effect of high-frequency alternating current transcutaneous stimulation over muscle strength: a controlled pilot study. J NeuroEngineering Rehabil 15, 103 (2018). https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-018-0443-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://0-doi-org.brum.beds.ac.uk/10.1186/s12984-018-0443-2

Keywords