Skip to main content

Table 2 Features extracted from the laboratory data

From: Automated detection of missteps during community ambulation in patients with Parkinson’s disease: a new approach for quantifying fall risk in the community setting

Feature

Axis

Hit

Specificity

Misstep

False alarm

Mean

Std

Mean

Std

Peak difference

Vertical [g]

93.10

83.31

0.34

0.18

0.60

0.20

Medio-lateral [g]

89.65

81.38

0.33

0.17

0.59

0.19

Anterior- posterior [g]

93.10

81.86

0.32

0.15

0.57

0.21

Yaw [deg/sec]

86.20

83.27

0.46

0.14

0.66

0.18

Pitch [deg/sec]

82.75

83.29

0.43

0.17

0.66

0.18

Roll [deg/sec]

89.65

82.81

0.53

0.21

0.66

0.20

Frequencies above threshold

FFT- Vertical

86.20

89.10

8.08

4.57

5.20

3.39

Entropy

2.43

0.42

2.09

0.57

Number of steps

Vertical

48.27

93.58

10.76

1.73

10.18

1.77

Maximum amplitude [g]

1.86

1.06

0.97

0.59

Number of steps

Anterior- posterior

86.20

91.77

8.68

1.86

8.85

1.91

Maximum amplitude [g]

1.89

1.07

0.85

0.64

Number of steps

Yaw

55.17

95.89

8.08

4.57

5.20

3.39

Maximum amplitude [deg/sec]

2.43

0.42

2.09

0.57

  1. The Hit and specificity for the first 6 features represent the performance of each feature on its own. For the other features, the results are the performance of each feature on its own but after the majority voting of the first 6.